| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 42605 units | |
| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 44755 units | |
| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 36727 units | |
| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 17925 units | |
| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 21591 units | |
| 2024-09-26 | Class II | No Marketing Application | Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | 31457 units | |
| 2019-01-04 | Class II | Packaging process control | Size 0 products were incorrectly packaged and labeled as a size 1 products. | 270 | 2020-04-20 |
| 2013-09-03 | Class III | Labeling mix-ups | Report that size 5 KLTD airways were packaged in pouches indicating a size of 2.5 and shipped to your facility. The correct size of the KLTD airway is clearly marked on the actual airway product and the physical size difference between 2.5 | 9 cases; 45 products | 2014-04-15 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2012-11-02 | Class II | Nonconforming Material/Component | The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that ha | 263 cases of product. (2,630 devices) | 2013-05-28 |
| 2011-01-24 | Not classified | Other | The firms contract manufacturer reported that the products were below the specification for hardness. This could contribute to a higher incidence of problems during the insertion and final positioning of the airway in the patient. In rare | 143 cases of 10 each case | 2012-06-29 |
| 2011-01-24 | Not classified | Other | The firms contract manufacturer reported that the products were below the specification for hardness. This could contribute to a higher incidence of problems during the insertion and final positioning of the airway in the patient. In rare | 138 cases 5 units per case, 690 units | 2012-06-29 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 airway kits total under recall | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 units total | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2011-01-14 | Not classified | Process control | King Systems has been notified by Triad Group, our supplier of sterile lubricating jelly, that a recall has been initiated for this product because the product may be “inadequately sterilized” and use of this product may result in “patient | 577,426 total units | 2013-02-20 |
| 2009-12-15 | Not classified | Labeling False and Misleading | King Systems was issued a warning letter from the Center for Devices and Radiological Health, Office of Compliance on October 28, 2009. The warning letter asserts that Kings Systems KLT(S)-D airway exceeds cleared indications in that it is | | 2012-08-16 |