Breezeway II® Plus Delivery Sheath
The Breezeway II Plus Delivery Sheath introduces diagnostic or therapeutic devices into the vasculature, including intracardiac, renal and peripheral locations. Options include an 11 French…
The Breezeway II Plus Delivery Sheath introduces diagnostic or therapeutic devices into the vasculature, including intracardiac, renal and peripheral locations. Options include an 11 French sheath with a 79 cm usable length and a fixed 180° curve. It is not intended for neural placement.
The assembly includes a braided sheath, a dilator, a hemostatic valve and a side port with a three-way stopcock. A radiopaque distal tip and side flush holes are built into the sheath. The dilator locks onto the sheath hub.
Listed Item
Delivery Sheath Plus, 11F 79CM 180° Lkit
Loader and Supply
The Plus version includes a loader that assists device insertion through the hemostatic valve. It accepts guidewires up to 0.038 inches and is supplied ethylene oxide sterile for single use by prescription.
Who Else Makes It
Concentrated in Jiangsu (5), Beijing (3), Shanghai (2). 2 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Mehow Innovative (Huizhou) Ltd. | Huizhou | Guangdong | Contract manufacturer | 11 |
| Tianjin Demax Medical Technology Co., Ltd. | Tianjin | Tianjin | Contract manufacturer | 5 |
| Steris Sterilization Technologies (Suzhou) Ltd | Suzhou | Jiangsu | Contract manufacturer | 44 |
| Shanghai INT Medical Instruments Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 10 |
| Lepu Medical Technology (Beijing) Co., Ltd. | Beijing | Beijing | Manufacturer | 7 |
| Beijing Target Medical Technologies Inc. | Beijing | Beijing | Contract manufacturer | 7 |
| Beijing Demax Medical Technology Co., Ltd. | Beijing | Beijing | Manufacturer | 5 |
| Curatia Medical Limited | Suzhou | Jiangsu | Specification developer | 5 |
| Tianjin Ethylene Oxide Sterilization Technology Co., Ltd. | Tianjin | Tianjin | Contract manufacturer | 4 |
| Shanghai Microport EP Medtech Co., Ltd. | Shanghai | Shanghai | Manufacturer | 3 |
| Becton Dickinson Medical Devices Co., Ltd. Suzhou | Suzhou | Jiangsu | Contract manufacturer | 2 |
| Cardiocycle Medical(Suzhou) Co., Ltd. | Suzhou | Jiangsu | Complaint-file establishment | 1 |
Showing 12 of 13. All 13 manufacturers in China registered for Introducer, Catheter
Record
| Listed by | Oscor Inc., United States |
|---|---|
| FDA product code | DYB |
| Device type | Introducer, Catheter |
| Device class | Class 2 |
| Regulation | 21 CFR 870.1340 |
| Establishment | Contract manufacturer, Contract steriliser, Repacker / relabeller, Complaint-file establishment, Manufacturer |
| 510(k) | K210627 |
| Record published | 2021-09-03 |
| Record revised | 2026-02-04 (version 3) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date, manufacturing date |
| Sizes | Outer diameter 2.9 to 5.7 mm; lumen/inner diameter 2.1 to 4.8 mm |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| BG140079045L | Sheath Overall Length 82 cm | In distribution |
| BG070110045L | Dilator Usable Length 115 cm | In distribution |
| BG070079045L | Lumen/Inner Diameter 2.6 mm | In distribution |
| BG110079180L | Outer Diameter 4.7 mm | In distribution |
| BG140079045L10 | Sheath Overall Length 82 cm | In distribution |
| BG100079045L | Dilator Usable Length 84 cm | In distribution |
| BG080079045L | Outer Diameter 3.6 mm | In distribution |
| BG120079045L | Dilator Overall Length 88 cm | In distribution |
| BG090079180L | Lumen/Inner Diameter 3.2 mm | In distribution |
| BG060110045L | Lumen/Inner Diameter 2.1 mm | In distribution |
| BG080110045L | Lumen/Inner Diameter 2.9 mm | In distribution |
| BG060079180L | Lumen/Inner Diameter 2.1 mm | In distribution |
| BG060079045L | Dilator Overall Length 87 cm | In distribution |
| BG100110045L | Lumen/Inner Diameter 3.5 mm | In distribution |
| BG140079045L08 | Sheath Usable Length 79 cm | In distribution |
| BG080079180L | Dilator Overall Length 87 cm | In distribution |
| BG090110045L | Lumen/Inner Diameter 3.2 mm | In distribution |
| BG090079045L | Dilator Overall Length 88 cm | In distribution |
| BG110079045L | Lumen/Inner Diameter 3.8 mm | In distribution |
| BG100079180L | Outer Diameter 4.3 mm | In distribution |
| BG070079180L | Outer Diameter 3.3 mm | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Demax | Beijing Demax Medical Technology Co., Ltd. | 2,841 |
| Arrow | Teleflex Medical LLC | 1,036 |
| Prelude IDeal™ | Merit Medical Systems, Inc. | 294 |
| Fast-Cath™ | St. Jude Medical | 213 |
| Flexor | Cook Incorporated | 182 |
| Prelude Ease™ | Merit Medical Systems, Inc. | 176 |
| Performer | Cook Incorporated | 174 |
| VSI | Teleflex Medical LLC | 154 |
| Peel-Away | Cook Incorporated | 153 |
| OptiSeal™ | Greatbatch Medical | 127 |
| Brilliant | Lepu Medical Technology (Beijing) Co., Ltd. | 122 |
| Super Sheath | Togo Medikit Co., Ltd. | 122 |
Showing 12 of 439. All products under Introducer, Catheter
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K210627 | 2021-06-03 |
| K151951 | 2015-09-22 |
| K140917 | 2014-07-09 |
| K140406 | 2014-04-17 |
| K130633 | 2014-02-13 |
| K122958 | 2012-12-13 |
| K122960 | 2012-12-13 |
| K122084 | 2012-08-29 |
| K120459 | 2012-06-29 |
| K101497 | 2010-10-20 |
| K090114 | 2009-06-04 |
| K073100 | 2007-11-30 |
| K890277 | 1989-03-10 |
| K831751 | 1983-10-20 |
Recalls
3 of these are Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2026-06-10 | Class I | Device Design | Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region. | 32,934 units | |
| 2026-06-10 | Class I | Device Design | Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region. | 355 units | |
| 2026-06-10 | Class I | Device Design | Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region. | 1332 units | |
| 2021-12-22 | Class II | Process control | There is potential for separation between the threaded hub and threaded cap at the proximal end of the device. | 549 units | |
| 2021-03-10 | Class II | Under Investigation by firm | Potential for open seal on sterile product | 500 devices | 2023-08-08 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 34 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 21 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 61 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 124 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 8 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 255 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 215 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 11 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 26 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 1,729 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 87 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 991 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 70 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 59 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 45 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 16 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 21 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 991 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 76 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 22 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 186 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 55 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 74 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 156 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 566 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 20 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 30 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 1,269 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 33 | 2019-07-26 |
| 2016-10-05 | Class II | Nonconforming Material/Component | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. | 110 | 2019-07-26 |
| 2008-07-03 | Not classified | Process design | Difficulty breaking the sheath hub and subsequently to peel the sheath off for Introducer Set, Adelante size 7F. | 396 | 2008-12-09 |
| 2005-09-13 | Not classified | Other | Sheath size is larger than dilator size causing insertion difficulties. | 6985 | 2012-07-08 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.