Detect®
Mapping Tool 10650 LES TST Detect 5P 18L
3manufacturers in China registered for this device type
2catalogue numbers in this listing
5510(k) clearances naming this product code
2recalls on this code
Who Else Makes It
Concentrated in Guangdong (2), Shandong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Dongguan King Medical Technology Co., Ltd. | Dongguan | Guangdong | Contract manufacturer | 14 |
| Shenzhen Launch Electrical Co., Ltd. | Shenzhen | Guangdong | Contract manufacturer | 8 |
| Bioxin Medical Co., Ltd. | Weihai | Shandong | Contract manufacturer | 8 |
Record
| Listed by | Medtronic, Inc., United States |
|---|---|
| FDA product code | LDF |
| Device type | Electrode, Pacemaker, Temporary |
| Device class | Class 2 |
| Regulation | 21 CFR 870.3680 |
| Establishment | Specification developer, Repacker / relabeller, Manufacturer, Exporter |
| 510(k) | K040812 |
| Record published | 2015-09-21 |
| Record revised | 2022-10-26 (version 6) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number |
| Storage | Storage environment temperature up to 40 °C |
| Packaging | Pack of 5 |
| Distribution end date | 2017-11-30 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 10650 | Withdrawn | |
| 10650 | Withdrawn |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| MYO/Wire® Ultra-Thin Temporary Pacing Wires | A & E Medical Corporation | 109 |
| Myo/Wire | A & E Medical Corporation | 76 |
| TB | Oscor Inc. | 40 |
| Temporary Myocardial Heartwire | Oscor Inc. | 27 |
| Pacel™ | St. Jude Medical Cardiovascular Division | 16 |
| Ethicon | Ethicon, Inc. | 15 |
| Flexon | Covidien LP | 13 |
| Electrode, Pacemaker, Temporary | Medtronic, Inc. | 12 |
| Nbih™ Temporary Pacing Electrode Catheter | C.R. Bard, Inc. | 11 |
| Electrode, Pacemaker, Temporary | C.R. Bard, Inc. | 10 |
| Streamline™ | Medtronic, Inc. | 10 |
| Steelex | B Braun Surgical SA | 9 |
Showing 12 of 31. All products under Electrode, Pacemaker, Temporary
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K242705 | 2025-01-17 |
| K190716 | 2019-08-08 |
| K171253 | 2017-05-25 |
| K161249 | 2016-08-19 |
| K140972 | 2014-06-06 |
Recalls
1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2016-06-16 | Class II | Nonconforming Material/Component | This Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage. | 11697 (5523 OUS, 6174 US) | 2018-02-13 |
| 2009-02-06 | Not classified | Process control | Medtronic is recalling one lot of STREAMLINE Bipolar Myocardial Temporary Pacing Leads, Model 6495, Lot JGD880205F. This lot of product contains incorrect inner pouch labeling, instead of 6495 it is labeled 6500. The outer carton, shelf b | 612 | 2011-05-07 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.