Pad-Pak
Combined Battery and Electrode Cartridge. Accessory to a therapeutic medical device in the range of Automated External Defibrillators.
Who Else Makes It
All of them in Guangdong. 3 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Amphenol Medical Technologies (Dongguan) Co., Ltd. | Dongguan | Guangdong | Contract manufacturer | 77 |
| Amphenol Alden Products Company | Shenzhen | Guangdong | Contract manufacturer | 24 |
| EMI (Dongguan) Asia Limited | Dongguan | Guangdong | Contract manufacturer | 9 |
Record
| Listed by | Heartsine Technologies, Ltd, United Kingdom |
|---|---|
| FDA product code | MKJ |
| Device type | Automated External Defibrillators (Non-Wearable) |
| Device class | Class 3 |
| Regulation | 21 CFR 870.5310 |
| Establishment | Specification developer, Manufacturer |
| 510(k) | P160008 |
| Record published | 2020-08-20 |
| Record revised | 2024-11-05 (version 3) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Single use | Yes |
| Prescription | Prescription only |
| Latex | Labelled latex-free |
| Label carries | lot or batch number, expiry date |
| Storage | Handling environment temperature 0 to 50 °C; Handling environment humidity 5 to 95 % RH; Storage environment humidity 5 to 95 % RH; Storage environment temperature 0 to 50 °C; Storage environment atmospheric pressure 57.2 to 106 KiloPascal; Handling environment atmospheric pressure 57.2 to 106 KiloPascal |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| Pad-Pak-07 | In distribution | |
| Pad-Pak-01 | In distribution | |
| Pad-Pak-01 | In distribution | |
| Pad-Pak-07 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| X Series | Zoll Medical Corporation | 862 |
| E Series | Zoll Medical Corporation | 571 |
| M Series | Zoll Medical Corporation | 219 |
| R Series | Zoll Medical Corporation | 216 |
| Lifepak® 15 monitor/defibrillator | Physio-Control, Inc. | 188 |
| AED Plus | Zoll Medical Corporation | 124 |
| Lifepak® 12 defibrillator/monitor | Physio-Control, Inc. | 122 |
| Lifepak CR Plus and Lifepak Express defibrillators | Physio-Control, Inc. | 111 |
| M Series Biphasic | Zoll Medical Corporation | 111 |
| PadPro | ConMed Corporation | 96 |
| R&D Batteries, Inc. | R & D Batteries, Inc. | 71 |
| Lifepak® 20 and Lifepak 20e defibrillator/monitors | Physio-Control, Inc. | 59 |
Showing 12 of 312. All products under Automated External Defibrillators (Non-Wearable)
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K142709 | 2015-03-27 |
| K123881 | 2013-07-11 |
| K052465 | 2006-12-12 |
| K042088 | 2004-09-02 |
| K041067 | 2004-05-25 |
| K023854 | 2003-02-14 |
Recalls
1 of these is Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2026-07-27 | Not classified | Process control | Due to a damaged (over stretched) membrane tail that causes an intermittent fault. | 11,912 | |
| 2025-09-18 | Class II | Under Investigation by firm | Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact. | 1,291,165 units | |
| 2025-06-30 | Class II | Process change control | Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks) | 190,516 units | |
| 2024-04-08 | Class II | Process control | Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue co | 183,248 | |
| 2024-02-14 | Class II | Process control | Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart | 12 | |
| 2023-04-26 | Class II | Process control | Single use battery and electrode cartridge may be rendered inoperable due to depleted battery cells which could prevent the defibrillator from analyzing patient condition or delivering therapy correctly. | 30,022 units | |
| 2023-02-01 | Class II | Process control | Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in therapy or no therapy. | 13 | |
| 2014-06-12 | Class II | Packaging process control | A small number of sealed foil pouches containing the electrodes were found to be difficult to open, resulting in potential delay in therapy. | 3245 | 2015-04-16 |
| 2012-09-13 | Class I | Device Design | Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on a | 81,982 in total | 2015-07-10 |
| 2005-10-05 | Not classified | Device Design | Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition. | 1929 units | 2012-02-29 |
| 2005-02-14 | Not classified | Nonconforming Material/Component | The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are interpreted by the AED so | 672 | 2011-06-08 |
PMA Approvals
A premarket approval is the full Class III route, granted on clinical evidence rather than on equivalence to an existing device. It is the highest regulatory bar a device of this type can clear.
| PMA number | Decision | Trade name | Approved to |
|---|---|---|---|
| P160008 | 2026-05-07 | HeartSines samaritan® Public Access Automated External Defibrillators (SAM 350P, SAM 360P and SAM 450P) and Accessories | Heartsine Technologies, Ltd |
| P170018 | 2025-11-06 | LIFEPAK® CR2 Automated External Defibrillator | Physio-Control, Inc. |
| P160032 | 2025-07-03 | Defibtech Lifeline; Defibtech Lifeline AUTO; Defibtech Lifeline VIEW; Defibtech Lifeline VIEW AUTO; Defibtech Lifeline E | Defibtech, LLC |
| P210015 | 2025-01-29 | Avive Automated External Defibrillator (AED) System | Avive Solutions Inc |
| P200005 | 2024-01-09 | TZ Medical Defibrillation Electrodes, Accessory Cables, and Connectors | TZ Medical, Inc. |
| P200007 | 2023-06-23 | HeartSync Multifunction Disposable Single-Use AED Defibrillator Pads | Graphic Controls D.B.A. Vermed |
| P200004 | 2022-06-16 | ConMed PadPro Multifunction Electrodes | ConMed Corporation |
| P160012 | 2021-07-08 | Infant/Child Reduced Energy Defibrillation Electrodes | Physio-Control, Inc. |
| P160033 | 2021-06-14 | Powerheart G5 AED | Zoll Medical Corporation |
| P190013 | 2021-02-02 | AED Battery Exchange (Models 9146-ABE, G5-ABE, 5070-ABE, FR3-ABE) | AED Battery Exchange, LLC |
| P180028 | 2020-08-22 | HeartStart FRx Defibrillator | Philips Medical Systems Technologies Ltd |
| P190007 | 2020-08-07 | Cardinal Health Multifunctional Defibrillation Electrode | Cardinalhealth |
| P160022 | 2020-06-11 | X Series and Propaq MD | Zoll Medical Corporation |
| P160015 | 2020-06-05 | AED 3 Defibrillator with Uni-Padz | Zoll Medical Corporation |
| P160028 | 2020-05-11 | Philips HeartStart FR3 Defibrillator | Philips Medical Systems Technologies Ltd |
| P160026 | 2020-02-13 | LIFEPAK Automated External Defibrillators (AEDs) | Physio-Control, Inc. |
| P160034 | 2019-08-20 | Powerheart® G3 Pro AED | Zoll Medical Corporation |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.