Vitros
VITROS Chemistry Products PROT Slides
Who Else Makes It
Concentrated in Zhejiang (1), Jilin (1), Shandong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Zhejiang Orient Gene Biotech | Huzhou | Zhejiang | Contract manufacturer | 86 |
| Dirui Industrial Co., Ltd. | Changchun | Jilin | Contract manufacturer | 84 |
| Biobase Biodustry(Shandong)Co., Ltd. | Jinan | Shandong | Manufacturer | 70 |
Record
| Listed by | Ortho-Clinical Diagnostics, Inc., United States |
|---|---|
| FDA product code | CEK |
| Device type | Biuret (Colorimetric), Total Protein |
| Device class | Class 2 |
| Regulation | 21 CFR 862.1635 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| Premarket notification | Exempt |
| Record published | 2016-08-26 |
| Record revised | 2023-06-21 (version 3) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date |
| Storage | Storage environment temperature -18 to -18 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 8208431 | In distribution | |
| 8392292 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Biobase | Biobase Biodustry(Shandong)Co., Ltd. | 17 |
| Synermed® | Infrared Laboratory Systems LLC | 10 |
| JAS Diagnostics | Drew Scientific, Ltd. | 6 |
| Pointe Scientific, Inc. | Horiba Instruments Incorporated | 5 |
| AU | Beckman Coulter, Inc. | 5 |
| Pointe Scientific, Inc. Total Protein | Horiba Instruments Incorporated | 5 |
| Synchron | Beckman Coulter, Inc. | 3 |
| Total Protein 2 (TP 2) | Randox Laboratories Limited | 3 |
| Total Protein | Roche Diagnostics Gmbh | 3 |
| Alinity | Abbott Ireland | 2 |
| Pointe Total Protein Reagent | Horiba Instruments Incorporated | 2 |
| Atellica CH Total Protein_2 | Siemens Healthcare Diagnostics Inc | 2 |
Showing 12 of 61. All products under Biuret (Colorimetric), Total Protein
Recalls Naming This Product
1 recall names this product by its own identifier, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Ortho-Clinical Diagnostics, Inc. | UDI-DI | 2020-12-15 | Class II | Device Design | Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides | US= 223,788 ; OUS:=168,120 Totlal: 491,908 | 2021-07-27 |
Recalls
2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2020-12-15 | Class II | Device Design | Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides | US=9802; OUS: 1298 Total: 11,100 | 2021-07-27 |
| 2013-06-12 | Class II | Equipment maintenance | Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues. | 19,438 units (8,845 units domestically & 10,593 units intern | 2017-04-11 |
| 2003-03-26 | Not classified | Other | Slides may exhibit random, elevated imprecision of results. | 29,273 (domestic); 4,967 (international) | 2004-11-16 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.