Pointe Glucose Reagent

Pointe Glucose Reagent packaged for use on the Mindray BA800. 8 x 66ml. For the in vitro quantitative measurement of glucose in serum.

Registry build 21.09.26
2manufacturers in China registered for this device type
1catalogue number
16recalls on this code

Who Else Makes It

Concentrated in Guangdong (1), Jiangsu (1).

Chinese manufacturers registered for FDA product code CFR
Manufacturer City Province Registered as Device types
Shenzhen Mindray BIO-Medical Electronics Co., Ltd. Shenzhen Guangdong Specification developer 61
Roche Diagnostics (Suzhou) Ltd. Suzhou Jiangsu Manufacturer 3

Record

Regulatory record for Pointe Glucose Reagent
Listed by Horiba Instruments Incorporated, United States
FDA product code CFR
Device type Hexokinase, Glucose
Device class Class 2
Regulation 21 CFR 862.1345
Establishment Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer
510(k) K961282
Record published 2021-01-30
Record revised 2025-10-02 (version 3)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Supplied as a kit Yes
Label carries lot or batch number, expiry date

Catalogue Numbers

Catalogue numbers for Pointe Glucose Reagent
Catalogue number Size Status
16-G7517-528 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Gluc-HK Randox Laboratories Limited 6
Glucoquant GlucoseHK Roche Diagnostics Gmbh 3
Glucose HK Roche Diagnostics Gmbh 3
AU Beckman Coulter, Inc. 3
Teco Diagnostics Glucose (Hexokinase) Reagent Set Teco Diagnostics 2
Yumizen C1200 Glucose HK Horiba ABX SAS 2
Architect Abbott Laboratories 2
Alinity Abbott Medical 2
Teco Diagnostics Glucose (Hexokinase) Liquid Reagent Teco Diagnostics 2
Atellica CH GluH_3 Siemens Healthcare Diagnostics Inc 2
Pointe Scientific, Inc. Horiba Instruments Incorporated 2
Architect Abbott Medical 2

Showing 12 of 43. All products under Hexokinase, Glucose

Recalls

1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2016-01-11 Class II Under Investigation by firm N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with t 216 2016-06-07
2009-10-19 Not classified Process design There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL. 400 kits plus 59 liters and 15 vials. 2010-09-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 4 2009-11-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 8 2009-11-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 63 2009-11-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 9 2009-11-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 10 2009-11-27
2008-09-10 Not classified Process control The inability of the product to maintain stated performance specifications through the stated shelf life. 37 2009-11-27
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 13 kits 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 153 kits 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 10 kits 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 279 bottles 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 14 kits 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 61 bottles 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 425 bottles 2007-11-19
2006-06-07 Not classified Other Product does not meet performance specifications through its labeled expiration period. 61 bottles 2007-11-19

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer