| 2025-03-21 | Class II | Component design/selection | ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (w | 6275 analyzers | |
| 2024-05-31 | Class II | Nonconforming Material/Component | Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specifi | 27,127 analyzers | |
| 2022-05-18 | Class II | Software design | There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+. | 1489 analyzers | |
| 2021-04-14 | Class II | Device Design | Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed. | Total=4,135 analyzers world-wide (only 1277 analyzers within | |
| 2020-05-26 | Class II | Software design | Potential risk of patient mix-up on analyzers due to software issues. | 3,899 units | 2023-02-08 |
| 2020-05-26 | Class II | Software design | Potential risk of patient mix-up on analyzers due to software issues. | 4,615 units | 2023-02-08 |
| 2020-05-11 | Class II | Under Investigation by firm | Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accessi | 1981 | 2023-02-24 |
| 2020-05-11 | Class II | Under Investigation by firm | Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accessi | 9373 units | 2023-02-24 |
| 2020-03-10 | Class II | Software design | Analyzer’s barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample res | 3,656 units | 2022-10-31 |
| 2020-03-06 | Class II | Component design/selection | Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could | 2 units | |
| 2020-02-03 | Class II | Software design | The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. T | 3,968 units | 2022-09-20 |
| 2019-11-21 | Class II | Nonconforming Material/Component | The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted | 544 | 2024-11-08 |
| 2019-11-21 | Class II | Nonconforming Material/Component | The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted | 1233 units | 2024-11-08 |
| 2018-02-27 | Class II | Device Design | There is a potential problem relating to the AQURE System, versions 2.3.0 and 2.3.1, that may result in patient mix-up. | 2 | 2019-03-20 |