RAPIDPoint® 500e Blood Gas System
The RAPIDPoint® 500e Blood Gas System is a compact, bench-top analyzer designed for in vitro diagnostic testing and issuitable for professional use in a point-of-care or central laboratory…
The RAPIDPoint® 500e Blood Gas System is a compact, bench-top analyzer designed for in vitro diagnostic testing and issuitable for professional use in a point-of-care or central laboratory environment. This system measures the following: blood gases,electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples.Additionally, the RAPIDPoint 500e Blood Gas System measures pH in pleural fluid.
Who Else Makes It
Registered in Guangdong.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Edan Instruments, Inc. | Shenzhen | Guangdong | Contract manufacturer | 56 |
Record
| Listed by | Siemens Healthcare Diagnostics Inc, United States |
|---|---|
| FDA product code | JFP |
| Device type | Electrode, Ion Specific, Calcium |
| Device class | Class 2 |
| Regulation | 21 CFR 862.1145 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| 510(k) | K192240 |
| Record published | 2020-04-15 |
| Record revised | 2021-11-19 (version 3) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | manufacturing date, serial number |
| Storage | Storage environment temperature 4 to 40 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 11416751 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Stat Profile | Nova Biomedical Corporation | 9 |
| Nova | Nova Biomedical Corporation | 8 |
| Calcium Electrode | Medica Corp. | 2 |
| SC90 | Radiometer Medical APS | 2 |
| pH/pCO2/pO2/Na/K/Ca Analyzer | Medica Corp. | 2 |
| Na/K/Cl/Ca/Li Solution pack | Medica Corp. | 1 |
| Rinse Solution S4980 REF 944-130 | Radiometer Medical APS | 1 |
| Nova/Stat Profile | Nova Biomedical Corporation | 1 |
| Rapidpoint® 350 Sampler Module | Siemens Healthcare Diagnostics Inc | 1 |
| RAPIDLab® 1200 pH Electrode Fill | Siemens Healthcare Diagnostics Inc | 1 |
| Roche 9180 Electrolyte Analyzer | Roche Diagnostics Gmbh | 1 |
| E-Ca Cassette Carton | Opti Medical Systems Inc. | 1 |
Showing 12 of 33. All products under Electrode, Ion Specific, Calcium
Recalls Naming This Product
4 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Siemens Healthcare Diagnostics Inc | 510(k) number | 2024-05-17 | Class II | Software Design Change | Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in th | 9 units US | |
| Siemens Healthcare Diagnostics Inc | 510(k) number | 2023-09-26 | Class II | Under Investigation by firm | There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels wi | 8,749 units | |
| Siemens Healthcare Diagnostics Inc | 510(k) number | 2022-06-01 | Class II | Under Investigation by firm | Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support. | 263 Boxes | |
| Siemens Healthcare Diagnostics Inc | 510(k) number | 2020-05-27 | Class II | Software design | The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes. | 436 | 2021-01-12 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.