Advia Centaur® Prge

Progesterone - CENTAUR - RGT - 50 tests

Registry build 21.09.26
4manufacturers in China registered for this device type
1catalogue number
9recalls on this code

Who Else Makes It

Concentrated in Hebei (1), Guangdong (1), Jiangsu (1).

Chinese manufacturers registered for FDA product code JLS
Manufacturer City Province Registered as Device types
Shijiazhuang Hipro Biotechnology Co., Ltd. Shijiazhuang Hebei Manufacturer 22
Shenzhen Yhlo Biotech Co., Ltd. Shenzhen Guangdong Manufacturer 19
Getein Biotech, Inc. Nanjing Jiangsu Manufacturer 19
Beijing Hotgen Biotech Co., Ltd. Beijing Beijing Manufacturer 13

Record

Regulatory record for Advia Centaur® Prge
Listed by Siemens Healthcare Diagnostics Inc, United States
FDA product code JLS
Device type Radioimmunoassay, Progesterone
Device class Class 1
Regulation 21 CFR 862.1620
Establishment Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter
510(k) K932955
Record published 2020-09-24
Record revised 2021-12-10 (version 3)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries lot or batch number, expiry date
Storage Storage environment temperature 2 to 8 °C; Special storage condition: Store Upright, Protect from Light, Do not freeze, Resuspend

Catalogue Numbers

Catalogue numbers for Advia Centaur® Prge
Catalogue number Size Status
10310305 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
MP Biomedicals LLC Mp Biomedicals, LLC 4
Elecsys Progesterone III Roche Diagnostics Gmbh 3
Alinity Abbott Ireland 3
Architect Abbott Ireland 3
Salimetrics Salimetrics LLC. 3
Atellica IM Prge Siemens Healthcare Diagnostics Inc 3
Salivary Progesterone Elisa DRG Instruments Gesellschaft mit beschränkter Haftung 2
Vitros Ortho-Clinical Diagnostics, Inc. 2
Immulite®/Immulite® 1000 Systems PRG Progesterone Siemens Healthcare Diagnostics Products GmbH 2
Salivary Progesterone HS Elisa DRG Instruments Gesellschaft mit beschränkter Haftung 2
Progesterone Luminescence Immunoassay IBL International Gmbh 2
Progesterone Elisa DRG Instruments Gesellschaft mit beschränkter Haftung 2

Showing 12 of 54. All products under Radioimmunoassay, Progesterone

Recalls Naming This Product

7 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Advia Centaur® Prge
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Siemens Healthcare Diagnostics Inc 510(k) number 2017-01-04 Class II Nonconforming Material/Component DHEA-S causes falsely elevated progesterone results. 76,107 2019-07-18
Siemens Healthcare Diagnostics Inc 510(k) number 2017-01-04 Class II Nonconforming Material/Component DHEA-S causes falsely elevated progesterone results. 1148 2019-07-18
Siemens Healthcare Diagnostics Inc 510(k) number 2017-01-04 Class II Nonconforming Material/Component DHEA-S causes falsely elevated progesterone results. 9076 2019-07-18
Siemens Healthcare Diagnostics Inc 510(k) number 2017-01-04 Class II Nonconforming Material/Component DHEA-S causes falsely elevated progesterone results. 11963 2019-07-18
Siemens Healthcare Diagnostics Inc 510(k) number 2015-01-27 Class II Process control ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL
(95.4 nmol/L)
9778 kits 2016-03-31
Siemens Healthcare Diagnostics Inc 510(k) number 2015-01-27 Class II Process control ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL
(95.4 nmol/L)
11980 kits 2016-03-31
Siemens Healthcare Diagnostics Inc 510(k) number 2015-01-27 Class II Process control ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL
(95.4 nmol/L)
1719 kits 2016-03-31

Recalls

2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2019-07-10 Class II Under Investigation by firm There is a potential for low discordant progesterone results on a subset of patient samples. A falsely low progesterone result may lead to the consideration for additional progesterone
supplementation.
6633 (4753 additional as of 1/13/20) 2024-05-28
2019-07-10 Class II Under Investigation by firm There is a potential for low discordant progesterone results on a subset of patient samples. A falsely low progesterone result may lead to the consideration for additional progesterone
supplementation.
4780 (6266 additional as of 1/13/20) 2024-05-28

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker