Advia Centaur® Shbg MCM

SHBG Master Curve Material Set - CENTAUR - CONS

Registry build 21.09.26
2manufacturers in China registered for this device type
1catalogue number
10510(k) clearances naming this product code
28recalls on this code

Who Else Makes It

Concentrated in Zhejiang (1), Jiangsu (1). One is registered as a contract manufacturer, able to build to a specification.

Chinese manufacturers registered for FDA product code JLW
Manufacturer City Province Registered as Device types
Hangzhou AllTest Biotech Co., Ltd. Hangzhou Zhejiang Contract manufacturer 48
Roche Diagnostics (Suzhou) Ltd. Suzhou Jiangsu Manufacturer 3

Record

Regulatory record for Advia Centaur® Shbg MCM
Listed by Siemens Healthcare Diagnostics Inc, United States
FDA product code JLW
Device type Radioimmunoassay, Thyroid-Stimulating Hormone
Device class Class 2
Regulation 21 CFR 862.1690
Establishment Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter
510(k) K083844
Record published 2016-09-24
Record revised 2021-11-18 (version 6)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries lot or batch number, expiry date
Storage Special storage condition: Store Upright; Storage environment temperature 2 to 8 °C
Distribution end date 2019-09-10

Catalogue Numbers

Catalogue numbers for Advia Centaur® Shbg MCM
Catalogue number Size Status
10283086 Withdrawn

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Atellica IM TSH3-UL Siemens Healthcare Diagnostics Inc 4
Elecsys TSH Roche Diagnostics Gmbh 4
Advia Centaur® TSH3-UL Siemens Healthcare Diagnostics Inc 3
Advia Centaur® TSH Siemens Healthcare Diagnostics Inc 3
Atellica IM TSH3-Ultra II Siemens Healthcare Diagnostics Inc 3
Free T3, Free T4, TSH Thyroid Panel AccuBind Vast Elisa Test System – 960 Wells Monobind, Inc. 3
Architect Abbott Ireland 2
Immulite 2000 Third Generation TSH Siemens Healthcare Diagnostics Products GmbH 2
Thyrotropin (TSH) Elisa DRG Instruments Gesellschaft mit beschränkter Haftung 2
AccuDiag ™ Diagnostic Automation, Inc. 2
Immulite® 2000 Systems TSH Third Generation TSH Siemens Healthcare Diagnostics Products GmbH 2
Maglumi 2000 TSH Shenzhen NEW Industries Biomedical Engineering Co., Ltd. 2

Showing 12 of 87. All products under Radioimmunoassay, Thyroid-Stimulating Hormone

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K251543 2026-02-06
K243570 2025-04-25
K233050 2024-04-04
K222116 2023-07-13
K160647 2016-05-31
K151792 2015-11-20
K150403 2015-05-01
K141999 2015-04-22
K083844 2009-05-08
K081074 2008-05-23

Recalls Naming This Product

3 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Advia Centaur® Shbg MCM
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Siemens Healthcare Diagnostics Inc 510(k) number 2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 22,770 kits 2016-05-27
Siemens Healthcare Diagnostics Inc 510(k) number 2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 155,345 kits 2016-05-27
Siemens Healthcare Diagnostics Inc 510(k) number 2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 172913 2016-05-10

Recalls

21 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2022-04-13 Class II Under Investigation by firm Lots incompatible with Test Definition (TDef) Version 1.4,
potentially lead to the generation of erroneous total T3 results
1,566
2022-04-13 Class II Under Investigation by firm Lots incompatible with Test Definition (TDef) Version 1.4,
potentially lead to the generation of erroneous total T3 results
2,321
2021-05-13 Class II Component design/selection Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject t US: 233 OUS: 233
2021-05-13 Class II Component design/selection Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject t US: 123 OUS: 751
2020-10-26 Class II Under Investigation by firm Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version tha US 392 units; OUS 449 units 2024-04-19
2020-10-26 Class II Under Investigation by firm Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some o US: 214 units; OUS: 1082 units 2024-04-19
2019-03-07 Class II Under Investigation by firm There were multiple issues identified in the system software which required an update. 366 2020-08-20
2019-03-07 Class II Under Investigation by firm There were multiple issues identified in the system software which required an update. 277 2020-08-20
2018-01-02 Class II Error in labeling The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly st 2023-02-10
2018-01-02 Class II Error in labeling The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly st 2023-02-10
2017-12-18 Class II Device Design In certain scenarios the ADVIA Centaur CP, XP and XPT immunoassay analyzers may exhibit reduced dilution recovery or become inoperable when used with Multi-Diluent 15 and certain lots of BNP and TSH3RUL Assays. 74,215
2017-03-21 Class II Under Investigation by firm The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, a 19, 927.00 2018-08-03
2016-05-26 Class II Under Investigation by firm Eight (8) issues were identified which may affect the results generated by the system software version. 446 units Total (3 domestically & 443 internationally) 2018-06-07
2016-04-28 Class II Under Investigation by firm The ADVIA Centaur XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any 292 units Total (3 domestically & 289 internationally) 2018-06-07
2015-04-20 Class II Nonconforming Material/Component Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well. 2256 2016-12-13
2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 8,052 kits 2016-05-27
2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 18584 2016-05-10
2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 33506 2016-05-10
2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 51792 2016-05-10
2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 39763 2016-05-10
2013-05-10 Class II Device Design A rare variant of TSH, identified in a small cluster of patients, is not detected. 13462 2016-05-10
2012-11-28 Class III Nonconforming Material/Component The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 Lot 831=19 units (all foreign consignees) and Lot 832=2 unit 2013-10-28
2012-11-28 Class III Nonconforming Material/Component The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 Lot 969=33 units (29 units foreign) Lot 970=7 units (42 unit 2013-10-28
2012-11-28 Class III Nonconforming Material/Component The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 Lot 742= 8 units all foreign consignees) Lot 743=7 units (al 2013-10-28
2010-04-12 Not classified Other Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results. 4187 2013-03-19

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer