Vitros
VITROS Immunodiagnostic Products TSH Reagent Pack
Who Else Makes It
Concentrated in Zhejiang (1), Jiangsu (1). One is registered as a contract manufacturer, able to build to a specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Hangzhou AllTest Biotech Co., Ltd. | Hangzhou | Zhejiang | Contract manufacturer | 48 |
| Roche Diagnostics (Suzhou) Ltd. | Suzhou | Jiangsu | Manufacturer | 3 |
Record
| Listed by | Ortho-Clinical Diagnostics, Inc., United States |
|---|---|
| FDA product code | JLW |
| Device type | Radioimmunoassay, Thyroid-Stimulating Hormone |
| Device class | Class 2 |
| Regulation | 21 CFR 862.1690 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| Record published | 2016-08-26 |
| Record revised | 2025-02-19 (version 4) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date, manufacturing date |
| Storage | Special storage condition: Do not freeze; Storage environment temperature 2 to 8 °C |
| Packaging | Case of 30 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 1912997 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Atellica IM TSH3-UL | Siemens Healthcare Diagnostics Inc | 4 |
| Elecsys TSH | Roche Diagnostics Gmbh | 4 |
| Advia Centaur® TSH3-UL | Siemens Healthcare Diagnostics Inc | 3 |
| Advia Centaur® TSH | Siemens Healthcare Diagnostics Inc | 3 |
| Atellica IM TSH3-Ultra II | Siemens Healthcare Diagnostics Inc | 3 |
| Free T3, Free T4, TSH Thyroid Panel AccuBind Vast Elisa Test System – 960 Wells | Monobind, Inc. | 3 |
| Architect | Abbott Ireland | 2 |
| Immulite 2000 Third Generation TSH | Siemens Healthcare Diagnostics Products GmbH | 2 |
| Thyrotropin (TSH) Elisa | DRG Instruments Gesellschaft mit beschränkter Haftung | 2 |
| AccuDiag ™ | Diagnostic Automation, Inc. | 2 |
| Immulite® 2000 Systems TSH Third Generation TSH | Siemens Healthcare Diagnostics Products GmbH | 2 |
| Maglumi 2000 TSH | Shenzhen NEW Industries Biomedical Engineering Co., Ltd. | 2 |
Showing 12 of 87. All products under Radioimmunoassay, Thyroid-Stimulating Hormone
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K083173 | 2008-12-22 |
| K081543 | 2008-10-23 |
Recalls
3 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2018-08-15 | Class II | Software design | Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result. | U.S.= 66 ex-U.S.= 76 | 2022-01-10 |
| 2018-08-15 | Class II | Software design | Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result. | U.S.= 1255 ex-U.S.= 1504 | 2022-01-10 |
| 2018-07-12 | Class II | Device Design | The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are | 68230 | |
| 2016-07-28 | Class III | Under Investigation by firm | A field correction was issued by the firm for VITROS TSH reagent lot 5040 due to a higher than expected frequency of customer calibration curves falling outside of the calibration quality parameters and therefore unusable to process samples | Domestic: 2883, Foreign: 1261 | 2018-06-11 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.