Vitros
VITROS Chemistry Products LAC Slides
Who Else Makes It
Concentrated in Shandong (1), Hunan (1), Jiangsu (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Biobase Biodustry(Shandong)Co., Ltd. | Jinan | Shandong | Manufacturer | 70 |
| AVE Science & Technology Co., Ltd. | Changsha | Hunan | Manufacturer | 22 |
| Eaglenos Sciences, Inc. | Nanjing | Jiangsu | Manufacturer | 4 |
Record
| Listed by | Ortho-Clinical Diagnostics, Inc., United States |
|---|---|
| FDA product code | KHP |
| Device type | Acid, Lactic, Enzymatic Method |
| Device class | Class 1 |
| Regulation | 21 CFR 862.1450 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| Premarket notification | Exempt |
| Record published | 2020-04-18 |
| Record revised | 2023-06-21 (version 2) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date |
| Storage | Storage environment temperature -18 to -18 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 8150112 | In distribution | |
| 8433880 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| StatSensor/StatStrip | Nova Biomedical Corporation | 5 |
| Stat Profile | Nova Biomedical Corporation | 5 |
| Lactate Gen.2 | Roche Diagnostics Gmbh | 4 |
| Lactate Plus | Nova Biomedical Corporation | 4 |
| Biobase | Biobase Biodustry(Shandong)Co., Ltd. | 3 |
| i-STAT | Abbott Point OF Care Inc. | 3 |
| Alinity | Abbott Medical | 2 |
| Fora | Fora Care Inc. | 2 |
| LA Cartridge | LifeHealth, LLC | 2 |
| L-Lactate (LAC) | Randox Laboratories Limited | 2 |
| ABX Pentra Lactic Acid | Horiba ABX SAS | 1 |
| Alinity | Abbott Ireland | 1 |
Showing 12 of 43. All products under Acid, Lactic, Enzymatic Method
Recalls Naming This Product
1 recall names this product by its own identifier, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|---|---|
| Ortho-Clinical Diagnostics, Inc. | UDI-DI | 2025-07-29 | Class II | Device Design | VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, ca | 2,320 units |
Recalls
2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2025-07-29 | Class II | Device Design | VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, ca | 2,908 units | |
| 2016-08-11 | Class II | Under Investigation by firm | Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays. | US: 32,081 sales units; Foreign: 29,932 sales units | 2018-06-11 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.