Advia Centaur® VitD DIL

Vitamin D Diluent (2PK) - CENTAUR - CAL - 2x25mL

Registry build 21.09.26
1manufacturer
1catalogue number
4510(k) clearances naming this product code
8recalls on this code

Who Else Makes It

Registered in Guangdong.

Chinese manufacturers registered for FDA product code MRG
Manufacturer City Province Registered as Device types
Shenzhen NEW Industries Biomedical Engineering Co., Ltd. Shenzhen Guangdong Manufacturer 4

Record

Regulatory record for Advia Centaur® VitD DIL
Listed by Siemens Healthcare Diagnostics Inc, United States
FDA product code MRG
Device type System, Test, Vitamin D
Device class Class 2
Regulation 21 CFR 862.1825
Establishment Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter
510(k) K110586
Record published 2016-09-24
Record revised 2023-06-02 (version 6)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries lot or batch number, expiry date
Storage Special storage condition: store upright; Storage environment temperature 2 to 8 °C

Catalogue Numbers

Catalogue numbers for Advia Centaur® VitD DIL
Catalogue number Size Per pack Status
10494100 2 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
BioPlex 2200 Bio-Rad Laboratories Inc. 6
Access 25(OH) Vitamin D Total Beckman Coulter, Inc. 3
Maglumi 25-OH Vitamin D Shenzhen NEW Industries Biomedical Engineering Co., Ltd. 3
Architect Abbott Medical 3
Atellica™ IM VitD Siemens Healthcare Diagnostics Inc 2
1,25 Dihydroxy Vitamin D RIA Immunodiagnostic Systems Ltd. 2
Maglumi 2000 25-OH Vitamin D Shenzhen NEW Industries Biomedical Engineering Co., Ltd. 2
Alinity Abbott Ireland 2
Architect Abbott Ireland 2
Elecsys Vitamin D total II Roche Diagnostics Gmbh 2
IDS-iSYS 25 VitDS Immunodiagnostic Systems Ltd. 2
Elecsys Vitamin D total III Roche Diagnostics Gmbh 2

Showing 12 of 50. All products under System, Test, Vitamin D

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K200509 2020-05-29
K162298 2017-03-16
K133156 2014-07-24
K110586 2011-10-14

Recalls Naming This Product

6 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Advia Centaur® VitD DIL
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Siemens Healthcare Diagnostics Inc 510(k) number 2016-10-05 Class II Device Design Vitamin D Total- Change in Correlation Between Serum and Plasma Specimen Tubes- negative bias with plasma vs. serum samples 69781 kits 2019-08-29
Siemens Healthcare Diagnostics Inc 510(k) number 2016-10-05 Class II Device Design Vitamin D Total- Change in Correlation Between Serum and Plasma Specimen Tubes- negative bias with plasma vs. serum samples 37437 kits 2019-08-29
Siemens Healthcare Diagnostics Inc 510(k) number 2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 17,731 kits 2016-05-27
Siemens Healthcare Diagnostics Inc 510(k) number 2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 12,787 kits 2016-05-27
Siemens Healthcare Diagnostics Inc 510(k) number 2014-04-10 Class II Device Design Samples containing fluorescein may show interference with the ADVIA Centaur Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays. 40,266 kits 2016-05-27
Siemens Healthcare Diagnostics Inc 510(k) number 2012-02-16 Not classified Nonconforming Material/Component Automatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes
diluted patient samples to under recover by approximately 50%.
442 units 2013-10-17

Recalls

2 of these are Class II. Class II means a temporary or reversible health consequence. 2 are recorded with the root cause “No Marketing Application”: the device was distributed without the premarket submission its class requires, a compliance finding rather than a defect. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2013-03-13 Class II No Marketing Application Negative bias in patient samples for Vitamin D 4133 2015-03-05
2013-03-13 Class II No Marketing Application Negative bias in patient samples for Vitamin D 3871 2015-03-05

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker