Clinisys Laboratory Solution

The Clinisys Laboratory Solution (CLS) is a laboratory information management system that is used to manage laboratory work. The core platform is intended to be used by trained laboratory…

Registry build 21.09.26

The Clinisys Laboratory Solution (CLS) is a laboratory information management system that is used to manage laboratory work. The core platform is intended to be used by trained laboratory professionals.

7manufacturers in China registered for this device type
4catalogue numbers in this listing
40recalls on this code

Who Else Makes It

Concentrated in Zhejiang (2), Guangdong (2), Jiangsu (1). 2 of them are registered as contract manufacturers, able to build to a buyer’s own specification.

Chinese manufacturers registered for FDA product code JQP
Manufacturer City Province Registered as Device types
Acon Biotech (Hangzhou) Co., Ltd. Hangzhou Zhejiang Contract manufacturer 29
Dongguan BAO Kang Dian ZI KE JI Limited Dongguan Guangdong Contract manufacturer 24
Jiangsu NEW Pentastar Medical Products Co., Ltd. Changzhou Jiangsu Exporter 146
Zhejiang Gene Science Co., Ltd. Shaoxing Zhejiang Manufacturer 31
Foshan Zike Technology Co., Ltd. Foshan Guangdong Manufacturer 9
Tianjin Empecs Medical Device Co., Ltd. Tianjin Tianjin Contract manufacturer 8
Shanghai Bestcovered Limited Shanghai Shanghai Manufacturer 2

Record

Regulatory record for Clinisys Laboratory Solution
Listed by CliniSys, Inc, United States
FDA product code JQP
Device type Calculator/Data Processing Module, For Clinical Use
Device class Class 1
Regulation 21 CFR 862.2100
Establishment Manufacturer
Premarket notification Exempt
Record published 2026-07-15

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Label carries lot or batch number

Catalogue Numbers

Catalogue numbers for Clinisys Laboratory Solution
Catalogue number Size Status
2025.3 In distribution
2025.2 In distribution
2023.3 In distribution
2025.1 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
SoftPathDx SCC Soft Computer 69
SoftMolecular SCC Soft Computer 67
SoftFlowCytometry SCC Soft Computer 67
SoftBiochemistry SCC Soft Computer 67
SoftCytogenetics SCC Soft Computer 67
SoftMic SCC Soft Computer 65
SoftLab SCC Soft Computer 64
SoftTrack SCC Soft Computer 61
SoftHLA SCC Soft Computer 61
SoftGenomics SCC Soft Computer 51
SoftGenePortal SCC Soft Computer 48
SoftTotalQC SCC Soft Computer 28

Showing 12 of 140. All products under Calculator/Data Processing Module, For Clinical Use

Recalls

5 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2012-03-15 Class II Software design Sunquest is recalling Sunquest Laboratory versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1 because when the Windows-based merge application is used and the “Event merge across patients” option is used, a receiving system (fo 126 sites 2012-10-02
2011-12-05 Class II Software design The recall was initiated because Sunquest has confirmed that the Sunquest Laboratory versions 6.4 and later while using Microbiology Result Entry (MRE), the specimen ribbon containing the Accession Number (AN) and the Battery Name may be mi 127 sites 2012-08-27
2011-11-15 Class III Software design Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument 127 sites 2012-10-19
2011-11-03 Class II Software design Sunquest is recalling Sunquest Laboratory version 7.0 because when saving an Accession to a non-current Episode/Event, a User-Defined field can be incorrectly replaced by the current Episode/Event User-Defined data. 14 sites 2012-11-06
2011-08-12 Class III Process design In GUI inquiry, searching by ID (CID) may return the wrong patient demographic data if the accession number associated to the CID is no longer inactive. 47 sites 2012-09-04
2011-04-28 Class III Software design In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur:
1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner containe
103 sites 2012-09-04
2010-10-05 Class II Software design The recall was initiated because Sunquest has confirmed that the Sunquest Laboratory LabAccess Results Workstation (LARS) may file results to the wrong patient by incorrectly combining results. 151 sites 2012-08-24
2009-11-19 Class II Nonconforming Material/Component The recall was initiated because Sunquest Information System has confirmed reports of an Outbound Phlebotomy Point of Care (POC) interface is combining patient orders inappropriately. 9 units 2012-11-06
2009-04-13 Class III Software design Sunquest is recalling the Sunquest Encompass versions 2.4 and later because the specimen collect time is displaying and printing one hour later than the actual collect time only when Daylight Saving Time is active. 23 clients 2012-08-22
2008-08-01 Not classified Software design While performing in-house testing for the Sunquest Laboratory System and under certain conditions, there have been instances in which an error that occurred at the database level or originated in the operating system (Cache)was not communic 139 2011-11-28
2008-04-17 Not classified Component design/selection Sunquest has identified a problem with IGO interfaces when performing an Online File Cleanup (OFC) that can lead to the test results of one patient filing to another patient when the order codes are the same. Performing an OFC prior to all 419 clients had the affected code 2011-12-05
2005-09-29 Not classified Other Customer complaint during testing of module prior to production use. Software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000, the system willfile the number 100 instea 67 2006-02-10
2005-05-13 Not classified Other Software anomaly may rarely allow modification of previously reported specimen results. 258 2006-02-10
2005-05-10 Not classified Other When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more suceptibilities on the culture will be deleted. 21 2005-08-30
2003-09-25 Not classified Other Software Defect. When orders are modified before the Roche Cobas Integra query for a specimen, the sample ID locks. 485 2004-09-03
2003-08-12 Not classified Other Software defect. Clinical Laboratory results failing quality assurance were filed directly to a patient’s chart without review. 479 2004-03-15
2003-07-24 Not classified Other Software Design Defect. Under certain conditions, results are removed. 47 2004-04-23
2003-07-23 Not classified Other Software Defect. Data used in the diagnosis and or treatment of a patient’s illness are missing from the patient result report. 466 2004-04-23
2003-07-22 Not classified Other Software design deficiency. If the dilution factor is added to the same result twice, incorrect specimen values are reported. 171 2004-06-25
2003-07-21 Not classified Other Incorrect demographics display which can result in incorrect information being used in diagnosis or treatment of patients. 125 2004-11-15
2003-07-07 Not classified Other Software defect causes orders to be lost under certain conditions of use. A typographical error at the Accept/Modfy/Reject prompt will cause the order not to be processed by the system. 473 2004-09-23
2003-07-02 Not classified Other Patient files become mixed up due to a software defect. 1 2004-04-23
2003-06-23 Not classified Other Software defect.
When Quality Assurance failure warnings are missing from a patient’s report abnormal results could be used for diagnosis or treatment.
157 2004-04-23
2003-06-16 Not classified Other Software anomaly allows patient reports to contain incomplete data used for diagnosis or treatment. 315 2004-04-23
2003-06-10 Not classified Other Software anomally
Results with a global flag failure that autofile without review can cause abnormal results being communicated without associated abnormal designations.
467 2004-04-23
2003-06-02 Not classified Other Susceptibility comments are ommited from Microbiology cumulative reports 471 2004-04-23
2003-05-13 Not classified Other Two clients reported Quality Assurance dates could not be entered after 5/19/03. 466 2004-04-23
2003-04-14 Not classified Other Incorrect results could be filed. 193 2004-04-21
2003-04-01 Not classified Other Urgent care STAT reports are not printing in a timely fashion. 109 2004-04-21
2003-03-26 Not classified Other With SMART enabled, for an accession number with multiple tests, credits and additions, the system may delete and reorder resulted tests. 463 2004-04-21
2003-03-19 Not classified Other Abnormal results are flagged as normal. 1 2004-12-01
2003-02-28 Not classified Other Patient results do not update properly when modified using on-line entry function. 460 2004-06-02
2003-02-19 Not classified Other Software anomaly. The HIDE command used to suppress suspected individual microbiology observations does not work in GUI inquiry 464 2004-06-02
2003-02-07 Not classified Other Missing patient data on Location Chart Copy Cumulative Reports. 478 2004-06-02
2003-02-06 Not classified Other Two customers report that the organism identified on the patient’s lab file do not match the reported susceptibility battery. 465 2004-05-07
2003-01-31 Not classified Other Software anomaly. Patient results may be filed in the incorrect patient file 542 2004-07-14
2003-01-31 Not classified Other Software anomaly. Rejecting the selected Autofiling status may result in the change being accepted rather than rejected. 542 2004-05-07
2003-01-17 Not classified Other Software anomally. Graphical display omits results containing a less than (<), greater than (>) or percent (%) symbol. 520 2004-05-07
2002-12-20 Not classified Other An incorrect collection date and time may be accepted and filed to the patient file. 465 2004-12-01
2002-12-20 Not classified Other Software anomally. English text code does not translate. 26 2004-04-23

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.