Liberty

Liberty Cycler Set, single stay•safe® PIN patient connector, 10-foot patient line, 30-inch drain line

Registry build 21.09.26
1manufacturer
2catalogue numbers in this listing
6510(k) clearances naming this product code
14recalls on this code

Who Else Makes It

Registered in Jiangsu. One is registered as a contract manufacturer, able to build to a specification.

Chinese manufacturers registered for FDA product code FKX
Manufacturer City Province Registered as Device types
Viant Medical Suzhou EPZ Co., Ltd. Jiangsu Jiangsu Contract manufacturer 3

Record

Regulatory record for Liberty
Listed by Fresenius Medical Care Renal Therapies Group, LLC, United States
FDA product code FKX
Device type System, Peritoneal, Automatic Delivery
Device class Class 2
Regulation 21 CFR 876.5630
Establishment Specification developer, Complaint-file establishment, Manufacturer
510(k) K173718
Record published 2024-03-19

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Supplied sterile Yes
Single use Yes
Prescription Prescription only
Label carries lot or batch number, expiry date
Packaging Case of 14

Catalogue Numbers

Catalogue numbers for Liberty
Catalogue number Size Status
050-87215A In distribution
050-87216A In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Sharesource Connectivity Platform Baxter Healthcare Corporation 7
Peritoneal Dialysis Merit Medical Systems, Inc. 6
Liberty® Fresenius Medical Care 5
VersiPD Fresenius Medical Care 4
VersiPD Fresenius USA, Inc. 3
Liberty® PDx Fresenius Medical Care 3
Premier Fresenius Medical Care 2
Safe-Lock® Fresenius Medical Care 2
HomeChoice APD System Baxter Healthcare Corporation 2
LOW Recirculation Volume APD SET Baxter Healthcare Corporation 2
Sharesource Baxter Healthcare Corporation 2
HomeChoice PRO APD System Baxter Healthcare Corporation 2

Showing 12 of 38. All products under System, Peritoneal, Automatic Delivery

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K222318 2022-10-31
K212522 2022-04-15
K181108 2018-05-24
K173718 2018-05-01
K171652 2017-09-15
K141145 2014-07-30

Recalls Naming This Product

1 recall names this product by its own identifier, the 510(k) it is cleared under.

Recalls naming the identifiers of Liberty
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Fresenius Medical Care Renal Therapies Group, LLC 510(k) number 2019-05-29 Class II Use error Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination. 33732973 Cassettes; 18,674 Liberty Cyclers 2020-11-19

Recalls

7 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2023-10-24 Class III Under Investigation by firm Affected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic. 149 US
2021-04-28 Class II Environmental control Exposed to freezing temperatures due to refrigerated truck malfunctions 6 Cases 2022-08-08
2020-09-12 Class II Device Design The device may detect an incorrect Heater Bag volume which may lead to a ‘Supply Bag Line Blocked’ alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra f 9,907 devices
2018-01-24 Class II Software design The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or d 9293 2020-05-11
2015-12-15 Class II Device Design The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitonea 14,094 units 2019-10-07
2013-10-14 Class II Device Design Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis 16,600 2021-05-18
2013-06-03 Class II Unknown/Undetermined by firm Cassette may leak during set up or during peritoneal dialysis treatment 5480 units 2014-06-18
2012-10-12 Class II Process control There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis. 560 cases, totaling 5590 units 2020-12-30
2012-03-30 Not classified Software design Increased risk of intraperitoneal volume (IIPV), referred to as Overfill 10,146 units 2016-08-04
2010-09-03 Not classified Nonconforming Material/Component Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis. 1524 units 2023-02-16
2010-09-03 Not classified Nonconforming Material/Component Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis. 363,265 units 2023-02-16
2010-09-03 Not classified Nonconforming Material/Component Cycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis. 5799 units 2023-02-16
2006-03-22 Not classified Other Stay safe trigger body may not connect to the extension set and result in associated peritonitis 141,770 units 2006-07-11

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.