Micro G-Jet®
The AMT Micro G-Jet is a low-profile transgastric-jejunal feeding device for patients weighing more than six kilograms who need nutrition delivered beyond the stomach. It can deliver…
The AMT Micro G-Jet is a low-profile transgastric-jejunal feeding device for patients weighing more than six kilograms who need nutrition delivered beyond the stomach. It can deliver enteral nutrition into the distal duodenum or proximal jejunum while providing gastric drainage through a separate pathway.
An inflatable balloon and low-profile external bolster retain the tube in the gastric stoma. The green jejunal and white gastric ports have one-way valves opened by the appropriate AMT extension set. A third port controls balloon inflation and deflation.
Listed Item
14FR x 1.7 cm x 15 cm – AMT micro G-Jet® Low Profile Transgastric-Jejunal Feeding Device (ENFit®), MDR
Drainage and Construction
The gastric port supports gravity drainage or low intermittent suction. The tube uses silicone and polyurethane, with reinforced jejunal tubing and barium sulfate for radiographic visualization. It is supplied sterile for prescription, single use.
Who Else Makes It
Concentrated in Jiangsu (17), Zhejiang (4), Guangdong (2). 11 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Amsino Medical (Kunshan) Co., Ltd. | Kunshan | Jiangsu | Contract manufacturer | 53 |
| Jiangsu JianYu Health Medical Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 35 |
| Jabil Circuit (Shanghai) Ltd. | Shanghai | Shanghai | Contract manufacturer | 31 |
| Conod Medical Shuyang Co., Ltd. | Shuyang | Jiangsu | Contract manufacturer | 22 |
| Jiangsu Caina Medical Co., Ltd. | Jiangyin | Jiangsu | Contract manufacturer | 20 |
| Jiangsu Weikang Jiejing Medical Apparatus Co., Ltd. | Shuyang | Jiangsu | Contract manufacturer | 18 |
| Mcreat ( Guangzhou ) BIO-Tech Co., Ltd. | Guangzhou | Guangdong | Contract manufacturer | 14 |
| Primecare Medical (Huzhou) Co., Ltd. | Hu Zhou | Zhejiang | Contract manufacturer | 9 |
| Yangzhou Xingwei Medical Technology Co., Ltd. | Yangzhou | Jiangsu | Contract manufacturer | 5 |
| Tessy Plastics Shanghai Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 4 |
| Flexan (Suzhou) Precision Rubber Products Co., Ltd. | Suzhou | Jiangsu | Contract manufacturer | 1 |
| Huzhou Junda Industry Co., Ltd. | Anji | Zhejiang | Contract manufacturer | 129 |
Showing 12 of 25. All 25 manufacturers in China registered for Tubes, Gastrointestinal (And Accessories)
Record
| Listed by | Applied Medical Technology, Inc., United States |
|---|---|
| FDA product code | KNT |
| Device type | Tubes, Gastrointestinal (And Accessories) |
| Device class | Class 2 |
| Regulation | 21 CFR 876.5980 |
| Establishment | Manufacturer, Exporter |
| 510(k) | K183508 |
| Record published | 2024-10-11 |
| Record revised | 2022-04-08 (version 2) |
Specifications
| MRI safety | MR Conditional |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Latex | Labelled latex-free |
| Label carries | lot or batch number, expiry date, manufacturing date |
| Sizes | Catheter gauge 8 to 14 Fr; length 0.8 to 30 cm |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| MGJ-1417-30 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1417-15-I | Length 1.7 cm | In distribution |
| MGJ-1412-10 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-22-I | Length 1.0 cm | In distribution |
| MGJ-1425-30 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1408-15 | Length 15 cm | In distribution |
| MGJ-1425-30-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1415-15 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-10-I | Length 1.0 cm | In distribution |
| MGJ-1412-30 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1425-22 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1417-30-I | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-15 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1425-22-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1408-15 | Length 15 cm | In distribution |
| MGJ-1417-10 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1425-22-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1415-22-I | Length 1.5 cm | In distribution |
| MGJ-1417-10 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-22-I | Length 1.0 cm | In distribution |
| MGJ-1417-15-I | Length 1.7 cm | In distribution |
| MGJ-1412-15-I | Length 1.2 cm | In distribution |
| MGJ-1412-30-I | Catheter Gauge 14 Fr | In distribution |
| MGJ-1412-22 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1415-22-I | Length 1.5 cm | In distribution |
| MGJ-1412-10 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1415-15 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-10-I | Length 1.0 cm | In distribution |
| MGJ-1408-10-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1425-30-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1408-10-I | Catheter Gauge 8 Fr | In distribution |
| MGJ-1415-10-I | Length 1.5 cm | In distribution |
| MGJ-1425-30 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1425-22 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1412-15-I | Length 1.2 cm | In distribution |
| MGJ-1412-30-I | Catheter Gauge 14 Fr | In distribution |
| MGJ-1415-10-I | Length 1.5 cm | In distribution |
| MGJ-1412-30 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1417-30 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1417-30-I | Catheter Gauge 14 Fr | In distribution |
| MGJ-1412-22 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1410-15 | Catheter Gauge 8 Fr | In distribution |
| MGJ-1417-22 | Catheter Gauge 14 Fr | In distribution |
| MGJ-1417-22 | Catheter Gauge 14 Fr | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Avanos | Avanos Medical, Inc. | 1,169 |
| Halyard | Avanos Medical, Inc. | 1,004 |
| G-Jet® Transgastric-Jejunal Feeding Device | Applied Medical Technology, Inc. | 376 |
| MiniONE Balloon Button | Applied Medical Technology, Inc. | 373 |
| Mini ONE® Balloon Button | Applied Medical Technology, Inc. | 336 |
| Kangaroo | Cardinal Health | 326 |
| NutraGlide Nasal Feeding Tube | Applied Medical Technology, Inc. | 279 |
| AMT Mini™ Balloon Button | Applied Medical Technology, Inc. | 231 |
| Vygon | Vygon Corp. | 229 |
| NeoDevices, Inc. | Neodevices Inc | 123 |
| Corflo cuBBy Lpgd | Corpak Medsystems | 98 |
| Kangaroo | Cardinal Health 200, LLC | 98 |
Showing 12 of 457. All products under Tubes, Gastrointestinal (And Accessories)
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K222846 | 2023-12-18 |
| K202539 | 2020-10-29 |
| K183508 | 2019-03-29 |
| K182804 | 2019-03-01 |
| K180026 | 2018-06-01 |
| K170319 | 2017-10-06 |
| K161413 | 2017-02-09 |
| K142971 | 2015-05-26 |
| K123716 | 2013-04-19 |
| K110804 | 2011-10-19 |
| K073034 | 2008-03-14 |
| K043027 | 2004-12-02 |
| K042300 | 2004-10-25 |
| K030784 | 2004-04-01 |
| K012476 | 2001-08-31 |
| K971434 | 1998-08-17 |
| K980305 | 1998-04-16 |
| K972478 | 1998-02-09 |
| K973893 | 1997-12-19 |
| K971757 | 1997-11-07 |
| K961345 | 1996-09-17 |
| K945618 | 1995-10-12 |
| K946140 | 1995-08-28 |
| K923488 | 1995-07-20 |
| K923692 | 1993-05-07 |
| K920894 | 1992-04-28 |
Recalls
11 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2025-10-06 | Class II | Process control | The nasal feeding tube may be occluded | ||
| 2025-10-06 | Class II | Process control | The nasal feeding tube may be occluded | ||
| 2025-10-06 | Class II | Process control | The nasal feeding tube may be occluded | ||
| 2025-10-06 | Class II | Process control | The nasal feeding tube may be occluded | ||
| 2025-10-06 | Class II | Process control | The nasal feeding tube may be occluded | ||
| 2025-08-12 | Class II | Process control | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces. | 840 units | |
| 2025-08-12 | Class II | Process control | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces. | 200 units | |
| 2025-08-12 | Class II | Process control | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces. | 390 units | |
| 2025-08-12 | Class II | Process control | The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces. | 1780 units | |
| 2024-05-14 | Class II | Process control | Devices were labelled with the incorrect guidewire labelling. | 74 units | |
| 2017-01-19 | Class III | Employee error | Expired expiration dates are listed on the pouches. | 110 | 2017-12-15 |
| 2017-01-19 | Class III | Employee error | Expired expiration dates are listed on the pouches. | 110 | 2017-12-15 |
| 2016-01-26 | Class II | Employee error | The firm received complaints that the product is shorter in length than indicated on the label. | 20 units | 2016-10-27 |