| GE Medical Systems, LLC | 510(k) number | 2026-05-01 | Class II | Component design/selection | GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit | 19587 units | |
| GE Medical Systems, LLC | 510(k) number | 2026-05-01 | Class II | Component design/selection | GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit | 23430 units | |
| GE Medical Systems, LLC | 510(k) number | 2026-05-01 | Class II | Component design/selection | GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit | 265 units | |
| Datex-Ohmeda, Inc. | 510(k) number | 2024-09-03 | Class I | Use error | GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors b | | |
| Datex-Ohmeda, Inc. | 510(k) number | 2024-09-03 | Class I | Use error | GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors b | | |
| Datex-Ohmeda, Inc. | 510(k) number | 2023-06-06 | Class I | Component design/selection | Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were noti | 769 units | |
| Datex-Ohmeda, Inc. | UDI-DI | 2023-06-06 | Class I | Component design/selection | Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were noti | 224 units | |
| Datex-Ohmeda, Inc. | 510(k) number | 2023-05-19 | Class II | Nonconforming Material/Component | GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devic | 4,540 units | |
| Datex-Ohmeda, Inc. | 510(k) number | 2023-05-19 | Class II | Nonconforming Material/Component | GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devic | 3,053 units | |
| GE Medical Systems, LLC | 510(k) number | 2020-04-28 | Class II | Device Design | Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed. | 5,883 units total | 2024-09-23 |
| GE Medical Systems, LLC | 510(k) number | 2020-04-02 | Class II | Device Design | During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor. | 204 units. | 2024-05-06 |
| GE Medical Systems, LLC | 510(k) number | 2019-10-11 | Class I | Device Design | Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall. | 23,693 units | 2024-08-15 |
| GE Medical Systems, LLC | 510(k) number | 2019-10-11 | Class I | Device Design | Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall. | 7814 units | 2024-08-15 |
| GE Medical Systems, LLC | 510(k) number | 2019-07-26 | Class II | Nonconforming Material/Component | Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall. | 59 devices (50 US, 9 OUS) | |
| GE Medical Systems, LLC | 510(k) number | 2018-09-25 | Class II | Process design | Potential electric leakage affecting the user. | 1,724 devices total | 2020-07-13 |
| GE Medical Systems, LLC | 510(k) number | 2013-01-08 | Class I | Process control | Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure. | 679 units | 2013-07-10 |