| 2026-06-30 | Class I | Under Investigation by firm | Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position. | 24528 units | |
| 2025-08-08 | Class II | Under Investigation by firm | Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation. | 14400 units | |
| 2024-12-20 | Class II | Process control | Specific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable. | | |
| 2024-12-20 | Class II | Process control | Specific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable. | | |
| 2024-12-20 | Class II | Process control | Specific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable. | | |
| 2024-02-08 | Class II | Under Investigation by firm | Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Clearlink Non-DEHP solution set listed below due to potential leaks originating from the drip chamber. | 26,592 units | |
| 2022-08-09 | Class I | Process design | Firm noted an increase in customer reports of leaks. | 511,728 units | 2024-08-09 |
| 2020-12-21 | Class II | Process design | There were customer reports of separation between the tubing and Male Luer Lock Adaptor. | 66,800 units | 2024-07-11 |
| 2019-12-12 | Class II | Device Design | Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector. | 43,495,400 units | |
| 2019-12-12 | Class II | Device Design | Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector. | 271,014,984 units | |
| 2015-11-30 | Class II | Under Investigation by firm | Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike adapter or solution set. | 9600 units | 2016-10-30 |
| 2015-11-30 | Class II | Under Investigation by firm | Potential for dark loose particulate matter (400 microns or smaller) on the spike of the spike adapter or solution set. | 9264 units | 2016-10-30 |
| 2015-05-21 | Class II | Packaging | Potential for a leak at the tubing to luer bond. | 23,600 | 2017-05-10 |
| 2014-07-14 | Class II | Nonconforming Material/Component | Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, | 234196 | 2016-03-15 |
| 2014-07-14 | Class II | Nonconforming Material/Component | Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, | 11184 | 2016-03-15 |
| 2014-07-14 | Class II | Nonconforming Material/Component | Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, | 107232 | 2016-03-15 |
| 2013-09-04 | Class II | Under Investigation by firm | The firm is issuing a voluntary recall due to the potential for pinholes in the packaging. Pinholes are a breach in the sterile barrier and could lead to potential contamination. | 3,200 units | 2017-04-19 |
| 2013-08-30 | Class II | Packaging | Potential tears in the pouch of six lots of Vented Spike Adapter Product Code 2C0471 | 49,344 units | 2017-04-19 |
| 2013-03-28 | Class II | Packaging process control | Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. | A) Product Code 2N1191: 53,850; B) Product Code 2N1206: 101, | 2017-07-31 |
| 2013-03-28 | Class II | Packaging process control | Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. | A) Product Code 2N3371: 228,200 units | 2017-07-31 |
| 2013-03-28 | Class II | Packaging process control | Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. | A) Product Code 2N8371: 513,000 units | 2017-07-31 |
| 2013-03-28 | Class II | Packaging process control | Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. | A) Code 2N8220: 5,950; B) Code 2N8221: 6,150; C) Code 2N8222 | 2017-07-31 |
| 2012-12-14 | Class II | Unknown/Undetermined by firm | Some of the individual extension set packages were improperly sealed, compromising the sterility of the product. | 15,257 units | 2016-12-23 |
| 2012-09-07 | Class I | Device Design | Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette. | 2,265,453 units | 2017-12-29 |
| 2012-09-07 | Class I | Device Design | Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette. | 1,995,040 units | 2017-12-29 |
| 2012-09-07 | Class I | Device Design | Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette. | 488,504 units | 2017-12-29 |
| 2012-09-07 | Class I | Device Design | Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette. | 278,420 units | 2017-12-29 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 1,440 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 2,160 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 720 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 4,560 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 528 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 2,352 units | 2012-11-01 |
| 2011-12-06 | Not classified | Employee error | During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection. | 4,272 units | 2012-11-01 |
| 2011-07-21 | Not classified | Employee error | The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be incorrectly loaded i | 4,752 units | 2012-06-18 |
| 2011-07-06 | Not classified | Packaging | Some of the individual extension set packages were improperly sealed, compromising the sterility of the product. | 9,800 sets | 2012-04-18 |
| 2011-05-31 | Not classified | Employee error | The lot of Baxter Solution Set for Epidural Use is being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the | 13,920 units | 2012-06-18 |
| 2010-10-21 | Not classified | Other | If the manifolds are connected in series, there may be a risk of the manifolds loosening or disconnecting at the luer connection between them. | 752,965 units in US and all lots in Japan | 2011-10-16 |
| 2010-09-09 | Not classified | Package design/selection | During packaging testing, Baxter observed tears and pinholes in Chevron-style pouches for certain infusion disposable products. The pinholes and tears could compromise the sterile barrier properties of the pouch. | 413,750 sets | 2011-09-07 |
| 2010-09-09 | Not classified | Package design/selection | During packaging testing, Baxter observed tears and pinholes in Chevron-style pouches for certain infusion disposable products. The pinholes and tears could compromise the sterile barrier properties of the pouch. | 60,400 sets | 2011-09-07 |