| 2026-05-12 | Not classified | Process control | Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and | 163,500,199 | |
| 2026-05-12 | Not classified | Process control | Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, | 107,800 | |
| 2025-12-19 | Class II | Process control | Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leak | 14,684,607 items | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 289098 units | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 29811 units | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 987911 units | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 265500 units | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 211750 units | |
| 2025-08-06 | Class II | Under Investigation by firm | Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. | 15546 units | |
| 2022-08-01 | Class II | Process control | Due to a manufacturing defect which results in a visible gap between the connector’s top and bottom housing. | Total of all affected units = 5,795,989 units | |
| 2022-08-01 | Class II | Process control | Due to a manufacturing defect which results in a visible gap between the connector’s top and bottom housing. | Total of all affected units = 5,795,989 units | |
| 2022-08-01 | Class II | Process control | Due to a manufacturing defect which results in a visible gap between the connector’s top and bottom housing. | Total of all affected units = 5,795,989 units | |
| 2022-08-01 | Class II | Process control | Due to a manufacturing defect which results in a visible gap between the connector’s top and bottom housing. | Total of all affected units = 5,795,989 units | |
| 2021-03-16 | Class II | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. | | |
| 2021-03-16 | Class II | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. | | |
| 2021-03-16 | Class II | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. | | |
| 2021-03-16 | Class II | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. | | |
| 2021-03-16 | Class II | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. | | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 19850 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 1856578 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 217522 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 33900 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 37000 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 55000 | |
| 2020-08-31 | Class II | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie | 325 | |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |
| 2020-07-06 | Class II | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. | Total of all products (Listed #1 thru 101) = 304735 units | 2023-05-11 |