| CareFusion 303, Inc | 510(k) number | 2026-02-11 | Class II | Under Investigation by firm | Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units | 1,025,567 | |
| CareFusion 303, Inc | 510(k) number | 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 867,362 units | |
| CareFusion 303, Inc | 510(k) number | 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 133,727 units | |
| CareFusion 303, Inc | 510(k) number | 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 86,393 units | |
| Infusion Pump Repair | 510(k) number | 2021-03-26 | Class I | Component design/selection | Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may r | 15 | |
| The Biomed Guys | 510(k) number | 2021-03-25 | Class I | Under Investigation by firm | Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free fl | 22 units | |
| Pacific Medical Group Inc. | 510(k) number | 2021-03-24 | Class I | Component design/selection | Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more beze | 2452 | |
| CareFusion 303, Inc | 510(k) number | 2021-03-03 | Class I | Process design | Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion | 160300 | |
| Step-HAR Medical LLC | 510(k) number | 2021-02-26 | Class I | Component design/selection | Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may r | 2 | 2024-08-15 |
| Tenacore LLC | 510(k) number | 2021-02-25 | Class I | Component design/selection | Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may r | 474 | 2024-08-05 |
| Tenacore LLC | 510(k) number | 2021-02-25 | Class I | Component design/selection | Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may r | 1494 | 2024-08-05 |
| BioMedical Equipment Service Co (Bmes) | 510(k) number | 2021-02-25 | Class I | Component design/selection | Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may r | 718 potentially affected pumps | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-08-04 | Class I | Component design/selection | The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion. | 180,186 parts & devices | |
| CareFusion 303, Inc | 510(k) number | 2020-08-04 | Class I | Reprocessing Controls | The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk pati | 313740 | |
| CareFusion 303, Inc | 510(k) number | 2020-08-04 | Class I | Reprocessing Controls | The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk pati | 28346 | |
| CareFusion 303, Inc | 510(k) number | 2020-08-04 | Class I | Device Design | Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious i | 248,519 | |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 37141 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 824277 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 1332354 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 122423 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 83532 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 38982 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 7854 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 17767 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Device Design | Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame | 1093588 | 2024-04-23 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Nonconforming Material/Component | If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power los | 37141 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class I | Nonconforming Material/Component | If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power los | 824277 | 2024-04-22 |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 1442097 | |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 130612 | |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 89497 | |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 40671 | |
| CareFusion 303, Inc | 510(k) number | 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 8558 | |
| CareFusion 303, Inc | 510(k) number | 2020-02-04 | Class I | Device Design | 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion al | 870070 | |
| CareFusion 303, Inc | 510(k) number | 2018-04-13 | Class I | Component design/selection | The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. externa | 607618 | 2023-12-26 |