| 2026-02-11 | Class II | Under Investigation by firm | Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units | 1,025,567 | |
| 2025-11-06 | Class II | Under Investigation by firm | Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices w | 189 devices | |
| 2025-07-08 | Class I | Under Investigation by firm | Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume. | 33,162,680 units total | |
| 2025-01-23 | Class II | Under Investigation by firm | Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left | 242,098 systems | |
| 2024-08-15 | Class II | Process control | Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | 20145 | |
| 2024-08-15 | Class II | Process control | Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | 10460 | |
| 2024-08-15 | Class II | Process control | Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | 180 | |
| 2024-08-15 | Class II | Process control | Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | 192964 | |
| 2024-08-15 | Class II | Process control | Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | 125017 device (finished devices and kits) | |
| 2024-05-21 | Class II | Under Investigation by firm | Due to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server. | 168 software version | |
| 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 86,393 units | |
| 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 133,727 units | |
| 2023-09-15 | Class I | Under Investigation by firm | Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. | 867,362 units | |
| 2023-01-24 | Class II | Software Design Change | Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Use | 83,185 | |
| 2021-03-03 | Class I | Process design | Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion | 160300 | |
| 2020-08-04 | Class I | Device Design | Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious i | 248,519 | |
| 2020-08-04 | Class I | Component design/selection | The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion. | 180,186 parts & devices | |
| 2020-08-04 | Class I | Reprocessing Controls | The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk pati | 313740 | |
| 2020-08-04 | Class I | Reprocessing Controls | The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk pati | 28346 | |
| 2020-06-30 | Class I | Nonconforming Material/Component | If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power los | 824277 | 2024-04-22 |
| 2020-06-30 | Class I | Device Design | Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame | 1093588 | 2024-04-23 |
| 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 1442097 | |
| 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 89497 | |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 37141 | 2024-04-22 |
| 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 130612 | |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 83532 | 2024-04-22 |
| 2020-06-30 | Class I | Nonconforming Material/Component | If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power los | 37141 | 2024-04-22 |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 824277 | 2024-04-22 |
| 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 8558 | |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 122423 | 2024-04-22 |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 1332354 | 2024-04-22 |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 7854 | 2024-04-22 |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 38982 | 2024-04-22 |
| 2020-06-30 | Class II | Device Design | LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim. | 40671 | |
| 2020-06-30 | Class I | Device Design | Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an a | 17767 | 2024-04-22 |
| 2020-02-04 | Class I | Device Design | 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion al | 870070 | |
| 2020-02-04 | Class I | Device Design | 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion al | 83015 | |
| 2020-02-04 | Class I | Device Design | 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion al | 114835 | |