| 2024-08-07 | Class I | Component design/selection | Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Addit | 14 | |
| 2024-06-14 | Class II | Device Design | Perfusor Space Syringe Pump’s instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, an | 9943 | |
| 2024-06-14 | Class II | Device Design | Perfusor Space Syringe Pump’s instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, an | 4519 | |
| 2024-06-14 | Class II | Device Design | Perfusor Space Syringe Pump’s instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, an | 14949 | |
| 2024-06-14 | Class II | Device Design | Perfusor Space Syringe Pump’s instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, an | 930 | |
| 2024-06-14 | Class II | Device Design | Perfusor Space Syringe Pump’s instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, an | 1663 | |
| 2012-12-05 | Class II | Component design/selection | B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) | 14,161 | 2014-10-24 |
| 2012-11-14 | Class I | Software design | This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in which the Outlook ES Safety Infusion System may halt infusion but the “RUN” Light Emitting Diodes (L | 6,919 | 2015-03-06 |
| 2012-10-31 | Class II | Software design | This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removi | 1,970 pumps added for this expansion | 2016-02-03 |
| 2012-10-24 | Class II | Component design/selection | B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue applies specifically to | 28,427 | 2015-07-27 |
| 2011-09-07 | Not classified | Software design | B Braun Outlook ES Safety Infusion System “Run” Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the “Run” LEDs advance as if the pumps were infusing. | 6321 | 2015-05-04 |
| 2010-08-24 | Not classified | Component design/selection | Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operati | 2,107 Pumps (582 Pumps to Winchester) | 2014-05-29 |
| 2010-07-20 | Not classified | Software design | The Outlook ES Dose Guard Care Area can inadvertently be exited when the pump enters the KVO (Keep Vein Open) state. The same condition can occur when a sequence of HOLD-HOLD key strokes are performed on the pump key panel. | 11,525 pumps | 2015-01-07 |