ICU Medical, Inc.

GTIN for CH-19, 1/EA

Registry build 21.09.26
1manufacturer
1catalogue number
4510(k) clearances naming this product code
40recalls on this code

Who Else Makes It

Registered in Shandong.

Chinese manufacturers registered for FDA product code ONB
Manufacturer City Province Registered as Device types
Shandong Ande Healthcare Apparatus Co., Ltd. Zibo Shandong Manufacturer 7

Record

Regulatory record for ICU Medical, Inc.
Listed by ICU Medical, Inc, United States
FDA product code ONB
Device type Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System
Device class Class 2
Regulation 21 CFR 880.5440
Establishment Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer
510(k) K173477
Record published 2025-04-25

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Supplied sterile Yes
Single use Yes
Prescription Prescription only
Label carries lot or batch number
Packaging Case of 50

Catalogue Numbers

Catalogue numbers for ICU Medical, Inc.
Catalogue number Size Status
CH-19 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
ICU Medical ICU Medical, Inc 255
Equashield Equashield Medical Ltd 83
BD PhaSeal BD Switzerland Sàrl 23
Halo J & J Solutions, Inc. 14
ProSeal Cstd Epic International (Thailand) Co., Ltd. 12
BD PhaSeal Optima BD Switzerland Sàrl 11
Equashield Cstd Equashield Medical Ltd 6
Luer Lock Adaptor 2 Swivel A Equashield Medical Ltd 1
OnGuard®2 Cstd Luer Lock Adaptor Simplivia Healthcare Ltd. 1
OnGuard2 Syringe Adaptor, 5 Units Simplivia Healthcare Ltd. 1
OnGuard 2 Cstd Bag Adaptor LL with Ultrasite Simplivia Healthcare Ltd. 1
Tevadaptor Catheter Adaptor Simplivia Healthcare Ltd. 1

Showing 12 of 60. All products under Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K190157 2019-10-23
K173477 2018-08-31
K170110 2018-08-10
K131549 2013-07-26

Recalls Naming This Product

8 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of ICU Medical, Inc.
Recalled by Identified by Initiated FDA class Root cause Reason Quantity
ICU Medical, Inc 510(k) number 2026-05-12 Not classified Process control Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, 28,550
ICU Medical, Inc 510(k) number 2026-05-12 Not classified Process control Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and 6,105,468
ICU Medical, Inc 510(k) number 2021-03-16 Class II Component design/selection Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
ICU Medical, Inc 510(k) number 2021-03-16 Class II Component design/selection Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
ICU Medical, Inc 510(k) number 2021-03-16 Class II Component design/selection Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
ICU Medical, Inc 510(k) number 2021-03-16 Class II Component design/selection Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
ICU Medical, Inc 510(k) number 2021-03-16 Class II Component design/selection Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
ICU Medical, Inc 510(k) number 2020-08-31 Class II Process control Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patie 534764

Recalls

38 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity
2026-03-24 Class II Process control Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential f 20,506
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 11000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 2450 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 200 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 6450 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 150 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 8475 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 11000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 16850 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 133,255 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 200 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 6150 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 600 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 900 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 200 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 3350 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 200 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 80 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 400 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 1100 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 11200 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 3000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 2000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 1600 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 5700 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 2000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 875 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 134,700 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 4000 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 2100 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 202900 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 53,050 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 950 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 1700 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 7950 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 950 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 9004 units
2025-09-08 Class II Under Investigation by firm Port weld of drug transfer device may separate or break during use and potentially result in a leak. 134700 units

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.