Varian Cryocare Touch™ System

The Varian Cryocare Touch System freezes and ablates tissue using extreme cold delivered through Endocare cryoprobes. It supports open, minimally invasive and endoscopic surgical…

Registry build 21.09.26

The Varian Cryocare Touch System freezes and ablates tissue using extreme cold delivered through Endocare cryoprobes. It supports open, minimally invasive and endoscopic surgical procedures, with indications spanning specified tumors, skin lesions and other tissue treatments.

The system uses the Joule-Thomson effect with argon and helium gases. High-pressure argon circulated through a cryoprobe produces cooling, while helium provides active thawing. Cooling and warming occur at the distal end of the probe shaft, under software control from the console.

Monitoring and Connections

The console has a 15-inch touchscreen, eight cryoprobe connections and eight TempProbe connections. Separate temperature probes monitor tissue in or around the targeted freeze zone. Wheels with locks and side handles support movement of the console.

3manufacturers in China registered for this device type
1catalogue number
3510(k) clearances naming this product code
6recalls on this code

Who Else Makes It

Concentrated in Guangdong (1), Beijing (1), Tianjin (1).

Chinese manufacturers registered for FDA product code GEH
Manufacturer City Province Registered as Device types
Shenzhen Kesite Digital Technology Co., Ltd. Shenzhen Guangdong Contract manufacturer 4
Hygea Medical Technology Co., Ltd. Beijing Beijing Manufacturer 2
Medinux (Tianjin) Technologies Co., Ltd. Tianjin Tianjin Manufacturer 1

Record

Regulatory record for Varian Cryocare Touch™ System
Listed by Varian Medical Systems, Inc, United States
FDA product code GEH
Device type Unit, Cryosurgical, Accessories
Device class Class 2
Regulation 21 CFR 878.4350
Establishment Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer
510(k) K201588
Record published 2020-08-14
Record revised 2025-10-13 (version 4)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries manufacturing date, serial number

Catalogue Numbers

Catalogue numbers for Varian Cryocare Touch™ System
Catalogue number Size Status
CRYO-38T In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Frigitronics Coopersurgical, Inc. 42
CryoProbe™ H&O Equipments, Nv/Sa 35
Wallach Coopersurgical, Inc. 32
Premier Premier Dental Products Co. 30
Hygea Hygea Medical Technology Co., Ltd. 26
truFreeze CSA Medical 20
Freezpen® H&O Equipments, Nv/Sa 19
Leisegang Cryotip for LM-900 Coopersurgical, Inc. 15
Integra® Miltex® Integra LifeSciences Production Corporation 13
cryoICE Cryoablation System Atricure, Inc. 13
CryoThin Medinux (Tianjin) Technologies Co., Ltd. 12
Erbe Erbe Elektromedizin GmbH 12

Showing 12 of 334. All products under Unit, Cryosurgical, Accessories

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K230271 2023-03-28
K211149 2022-01-05
K201588 2020-08-11

Recalls

1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2012-12-20 Class II Incorrect or no expiration date The Endocare 3.8mm RenalCryo Cryoprobe is being recalled because the label on the outer box has the incorrect expiration date. 50 units 2013-04-22
2012-02-27 Not classified Labeling mix-ups HealthTronics has initiated this recall of these products because there have been instances where the temperature probes (aka TempProbes) in the kit may be mislabeled. In some instances, the label near the prove does not match the label at 9274 units 2013-03-19
2010-11-04 Not classified Other Mislabeled : The label on the outer box and inner pouch of the EndoCare 2.4mm CryoProbe (CRYO-44F) inaccurately reference the CVA2400 (2.4mm Variable Probe aka: V-Probe). 33 Units 2011-04-14
2008-12-05 Not classified Other Endocare received a report of a patient death that occurred during a prostate cryoablation procedure. An autopsy attributed the cause of death to an air embolism. The safety notice for the cryoprobes was issued to refocus attention on t 22,741 units total for all products 2010-10-21
2008-12-05 Not classified Other Endocare received a report of a patient death that occurred during a prostate cryoablation procedure. An autopsy attributed the cause of death to an air embolism. The safety notice for the cryoprobes was issued to refocus attention on th 22,741 units total for all products 2010-10-21
2002-10-25 Not classified Other Device design failure, connectivity. 933 2003-04-15

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer