VivAer® Stylus
The VivAer® Stylus is a disposable handheld device capable of delivering bipolar radiofrequency energy to tissue. The VivAer® Stylus consists of a handle, shaft and treatment tip. An array…
The VivAer® Stylus is a disposable handheld device capable of delivering bipolar radiofrequency energy to tissue. The VivAer® Stylus consists of a handle, shaft and treatment tip. An array of bipolar electrodes is positioned on a nonconductive tip which is attached to a handle via a nonconductive shaft. A temperature sensor is located on the tip to monitor tissue temperature. The Stylus is attached to a Temperature controlled radiofrequency generator (Aerin Console) via a flexible cable.
Who Else Makes It
Concentrated in Guangdong (24), Zhejiang (13), Shanghai (10). 23 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Amphenol Medical Technologies (Dongguan) Co., Ltd. | Dongguan | Guangdong | Contract manufacturer | 77 |
| Shandong Weigao Orthopaedic Device Co., Ltd. | Weihai | Shandong | Contract manufacturer | 42 |
| Hangzhou Kangji Medical Instrument Co., Ltd. | Hangzhou | Zhejiang | Contract manufacturer | 27 |
| Modern Medical Equipment Manufacturing Ltd. | Heng Li, Dongguan | Guangdong | Contract manufacturer | 26 |
| Amphenol Alden Products Company | Shenzhen | Guangdong | Contract manufacturer | 24 |
| Hangzhou Endo Medical Instrument Co., Ltd. | Hangzhou | Zhejiang | Contract manufacturer | 19 |
| Hong Qiangxing (Shen Zhen) Electronics Limited | Shenzhen | Guangdong | Contract manufacturer | 19 |
| Shenzhen MED-Link Electronics Tech Co., Ltd. | Shenzhen | Guangdong | Contract manufacturer | 18 |
| Sanbor Medical Xiamen, Inc. | Xiamen | Fujian | Contract manufacturer | 16 |
| ATL Technology DG,Ltd. | Dongguan | Guangdong | Contract manufacturer | 14 |
| Jiangyin Sinbon Electronics Co Ltd | Jiangyin | Jiangsu | Contract manufacturer | 13 |
| Guangzhou Huada Biotechnology Co., Ltd. | Guangzhou | Guangdong | Contract manufacturer | 9 |
Showing 12 of 65. All 65 manufacturers in China registered for Electrosurgical, Cutting & Coagulation & Accessories
Record
| Listed by | Aerin Medical Inc., United States |
|---|---|
| FDA product code | GEI |
| Device type | Electrosurgical, Cutting & Coagulation & Accessories |
| Device class | Class 2 |
| Regulation | 21 CFR 878.4400 |
| Establishment | Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer |
| 510(k) | K200300 |
| Record published | 2024-01-05 |
| Record revised | 2025-10-22 (version 4) |
Specifications
| MRI safety | MR Unsafe |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Label carries | lot or batch number, expiry date |
| Storage | Handling environment humidity 15 to 90 % RH; Handling environment temperature -18 to 60 °C |
| Distribution end date | 2025-10-01 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| FG1787 | In distribution | |
| FG722 | Withdrawn |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Surgical Direct | Surgical Direct, Inc. | 5,729 |
| Speedlock | National Advanced Endoscopy Devices, Inc. | 5,212 |
| Stingray Surgical Products | Stingray Surgical Products L.L.C. | 4,647 |
| Kirwan | Kirwan Surgical Products, LLC. | 1,901 |
| Millennium | Avalign Technologies, Inc. | 1,779 |
| DANmed | Dannoritzer Medizintechnik GmbH & Co. KG | 1,590 |
| Elmed | Elmed, Inc. | 1,519 |
| Symmetry Olsen | Aspen Surgical Products, Inc. | 1,219 |
| Guenter Bissinger Medizintechnik GmbH | Günter Bissinger Medizintechnik GmbH | 1,117 |
| Ambler Surgical | Ambler Surgical, LLC | 756 |
| V. Mueller | Steris Corporation | 661 |
| Aesculap | Aesculap AG | 544 |
Showing 12 of 2,922. All products under Electrosurgical, Cutting & Coagulation & Accessories
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K260522 | 2026-05-18 |
| K242630 | 2024-12-10 |
| K221907 | 2022-07-29 |
| K200300 | 2020-04-13 |
| K192471 | 2019-12-20 |
| K172529 | 2017-12-05 |
| K162810 | 2017-01-11 |
| K161994 | 2016-09-22 |
| K150637 | 2015-11-19 |
Recalls Naming This Product
1 recall names this product by its own identifier, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|---|---|
| Aerin Medical Inc. | 510(k) number | 2024-11-12 | Class II | Under Investigation by firm | Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console. | 479 units |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.