Origio Sequential Cleav
ORIGIO Sequential Cleav with phenol red. 10 ml
Who Else Makes It
Concentrated in Guangdong (3), Shandong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Dongguan King Medical Technology Co., Ltd. | Dongguan | Guangdong | Contract manufacturer | 14 |
| Shenzhen VitaVitro Biotech Co., Ltd. | Shenzhen | Guangdong | Manufacturer | 7 |
| Gimbo Medical Technology Shenzhen Co., Ltd. | Shenzhen | Guangdong | Manufacturer | 3 |
| Shandong Wego Ruisheng Medical Device Co., Ltd. | Weihai | Shandong | Manufacturer | 2 |
Record
| Listed by | Coopersurgical, Inc., United States |
|---|---|
| FDA product code | MQL |
| Device type | Media, Reproductive |
| Device class | Class 2 |
| Regulation | 21 CFR 884.6180 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer |
| 510(k) | K133912 |
| Record published | 2026-02-05 |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Supplied sterile | Yes |
| Prescription | Prescription only |
| Label carries | lot or batch number, expiry date |
| Storage | Special storage condition: Protect from light; Storage environment temperature 2 to 8 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 83030060M | In distribution | |
| 83030010M | In distribution | |
| 83040010M | In distribution | |
| 83040060M | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| LifeGlobal® Media | LifeGlobal Group, LLC | 93 |
| Sydney IVF | William A. Cook Australia PTY, Ltd. | 29 |
| LifeGlobal® Media | Coopersurgical, Inc. | 27 |
| VitaVitro | Shenzhen VitaVitro Biotech Co., Ltd. | 22 |
| Nidacon | Nidacon International AB | 19 |
| Sil-Select Plus Lower Layer | Fertipro NV | 8 |
| FertiCult Mineral Oil | Fertipro NV | 8 |
| Dewin | Donnevie Medical Technology (Shanghai) Co., Ltd. | 8 |
| SpermGrad | Vitrolife Sweden AB | 8 |
| Sequential Culture Media | Kitazato Corporation | 8 |
| Sperm Wash Medium | Invitrocare Inc | 7 |
| Enhance-S Plus | Vitrolife Inc. | 7 |
Showing 12 of 360. All products under Media, Reproductive
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K173731 | 2018-03-02 |
| K170560 | 2017-04-04 |
Recalls
8 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2024-02-16 | Class II | Under Investigation by firm | Low concentration of sodium pyruvate in the wash could compromise gamete development | 980 units | |
| 2024-02-16 | Class II | Under Investigation by firm | Low concentration of sodium pyruvate in the wash could compromise gamete development | 2680 units | |
| 2024-02-16 | Class II | Under Investigation by firm | Low concentration of sodium pyruvate in the wash could compromise gamete development | 770 units | |
| 2023-12-20 | Class II | Under Investigation by firm | The firm has become aware of a sudden increase in complaints for three associated lots. Performance issues may lead to impaired embryo development prior to the blastocyst stage. | 994 total (481 US; 513 OUS) | |
| 2023-02-13 | Class II | Labeling Change Control | It has come to CooperSurgical’s attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development. | 190 units | |
| 2022-12-15 | Class II | Labeling Change Control | The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo. | 489 units | |
| 2022-06-13 | Class II | Error in labeling | There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version. | 403 units | |
| 2020-09-08 | Class II | Under Investigation by firm | There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrificati | 531 total boxes | 2021-01-11 |
| 2016-08-25 | Class III | Packaging change control | Products contain international package insert and not the correct USA package insert | 1 unit | 2017-02-01 |
| 2016-08-25 | Class III | Packaging change control | Products contain international package insert and not the correct USA package insert | 4 units | 2017-02-01 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.