Visulux®
Catalogue number 000000-1100-155, listed by Carl Zeiss Meditec Inc, United States, under Biomicroscope, Slit-Lamp, Ac-Powered. 7 manufacturers in China are registered for this product code.
Who Else Makes It
Concentrated in Jiangsu (3), Chongqing (2), Beijing (1). One is registered as a contract manufacturer, able to build to a specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Chongqing Kanghua Ruiming S&T Co., Ltd. | Chongqing | Chongqing | Contract manufacturer | 3 |
| Beijing Hongdee Corporation | Beijing | Beijing | Private-label distributor | 43 |
| Taizhou City K&W International Trade Co., Ltd. | Jiangyan | Jiangsu | Repacker / relabeller | 38 |
| Lianyungang Tiannuo Optical Instrument Co., Ltd. | Lianyungang | Jiangsu | Contract manufacturer | 31 |
| 66 Vision Tech Co., Ltd. | Suzhou | Jiangsu | Private-label distributor | 24 |
| Shanghai Mediworks Precision Instruments Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 13 |
| Chongqing Sunkingdom Medical Instrument Co., Ltd. | Chongqing | Chongqing | Manufacturer | 11 |
Record
| Listed by | Carl Zeiss Meditec Inc, United States |
|---|---|
| FDA product code | HJO |
| Device type | Biomicroscope, Slit-Lamp, Ac-Powered |
| Device class | Class 2 |
| Regulation | 21 CFR 886.1850 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Manufacturer |
| 510(k) | K161607 |
| Record published | 2017-05-04 |
| Record revised | 2022-12-20 (version 5) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Label carries | manufacturing date, serial number |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 000000-1100-155 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| MediWorks | Shanghai Mediworks Precision Instruments Co., Ltd. | 30 |
| S4 Optik | Costruzioni Strumenti Oftalmici C.S.O. SRL | 24 |
| Marco | Takagi Seiko Co., Ltd. | 17 |
| Gilras | U.S. Ophthalmic, L.L.C. | 16 |
| Visionix | Luneau Technology Operations | 15 |
| LED Slit Lamp | Right MFG. Co., Ltd. | 13 |
| Tiannuo / Tnmed | Lianyungang Tiannuo Optical Instrument Co., Ltd. | 12 |
| Eli Ezer | U.S. Ophthalmic, L.L.C. | 12 |
| Inami | Inami & Co., Ltd. | 11 |
| Zoom Slit Lamp | Right MFG. Co., Ltd. | 8 |
| Slit Lamp BQ 900 | Haag-Streit AG | 8 |
| PolyLock Small Bone | Fusion Orthopedics, LLC | 7 |
Showing 12 of 168. All products under Biomicroscope, Slit-Lamp, Ac-Powered
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K173771 | 2018-08-24 |
| K162684 | 2017-05-31 |
| K170171 | 2017-04-06 |
| K161607 | 2017-02-23 |
| K143275 | 2015-07-10 |
| K141068 | 2014-09-09 |
| K133476 | 2014-02-12 |
| K122418 | 2013-04-12 |
| K101182 | 2010-11-03 |
Recalls
2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2013-10-23 | Class II | Software in the Use Environment | The user database delivered for download contained an erroneous value. Specifically the “aO” constant of the lens AMO Sensar AR40E is listed as “0.0” in the internal lens database of the IOLMaster. The correct value is -2.420. This could le | 4,907 units | 2014-04-17 |
| 2013-10-23 | Class II | Software in the Use Environment | The user database delivered for download contained an erroneous value. Specifically the “aO” constant of the lens AMO Sensar AR40E is listed as “0.0” in the internal lens database of the IOLMaster. The correct value is -2.420. This could le | 1,201 units | 2014-04-17 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.