Inserter Shaft, Flex-Thread™
Catalogue number 8435-1, listed by Flower Orthopedics Corporation, United States, under Rod, Fixation, Intramedullary And Accessories. 17 manufacturers in China are registered for this product code.
Who Else Makes It
Concentrated in Jiangsu (7), Shanghai (3), Zhejiang (3). 10 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Zhejiang Biomet Medical Products Co., Ltd. | Jinhua | Zhejiang | Contract manufacturer | 74 |
| Apex (Guangzhou) Tools & Orthopedics Company | Guangzhou | Guangdong | Contract manufacturer | 62 |
| Changzhou Biomet Medical Devices Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 54 |
| Paragon Medical Device (Changzhou) Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 53 |
| Canwell Medical Co., Ltd. | Jinhua | Zhejiang | Contract manufacturer | 42 |
| LX Precision Shanghai Co Ltd | Shanghai | Shanghai | Contract manufacturer | 19 |
| Stryker (Suzhou) Medical Technology Co., Ltd. | Suzhou | Jiangsu | Contract manufacturer | 19 |
| Suzhou Ahno Medical Devices Co., Ltd. | Suzhou | Jiangsu | Contract manufacturer | 17 |
| Turnxon Precision Co., Ltd. | Dongguan | Guangdong | Contract manufacturer | 17 |
| Suzhou Kangli Orthopaedics Instrument Co., Ltd. | Zhangjiagang | Jiangsu | Contract manufacturer | 11 |
| Zimmer (Shanghai) Medical International Trading Co., Ltd. | Shanghai | Shanghai | Exporter | 106 |
| Shanghai JPY ION-Tech Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 70 |
Showing 12 of 17. All 17 manufacturers in China registered for Rod, Fixation, Intramedullary And Accessories
Record
| Listed by | Flower Orthopedics Corporation, United States |
|---|---|
| FDA product code | HSB |
| Device type | Rod, Fixation, Intramedullary And Accessories |
| Device class | Class 2 |
| Regulation | 21 CFR 888.3020 |
| Record published | 2021-03-19 |
| Record revised | 2024-02-12 (version 2) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Sterilisation | Moist Heat or Steam Sterilization |
| Prescription | Prescription only |
| Label carries | lot or batch number |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 8435-1 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| T2 | Stryker GmbH | 1,373 |
| Autobahn | Globus Medical, Inc. | 1,258 |
| Gamma | Stryker GmbH | 1,120 |
| Summa Orthopaedics Femur Nail System | Jeil Medical Corporation | 742 |
| Arix Femur Nail System | Jeil Medical Corporation | 742 |
| Orthofix | Orthofix SRL | 721 |
| Rod, Fixation, Intramedullary And Accessories | Synthes Gmbh | 715 |
| Trigen | Smith & Nephew, Inc. | 681 |
| Arthrex® | Arthrex, Inc. | 665 |
| Precice® Intramedullary Limb Lengthening System | NuVasive Specialized Orthopedics, Inc. | 596 |
| Apex | Orthoxel Designated Activity Company | 569 |
| Natural Nail® | Zimmer, Inc. | 552 |
Showing 12 of 2,207. All products under Rod, Fixation, Intramedullary And Accessories
Recalls Naming This Product
1 recall names this product by its own identifier, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Flower Orthopedics Corporation DBA Conventus Flower | UDI-DI | 2021-11-02 | Class II | Device Design | Mating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available | 48 units | 2023-10-12 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.