CHS
Intertan Compression Screw 30MM
Who Else Makes It
Concentrated in Shanghai (2), Jiangsu (1). One is registered as a contract manufacturer, able to build to a specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| LX Precision Shanghai Co Ltd | Shanghai | Shanghai | Contract manufacturer | 19 |
| Zimmer (Shanghai) Medical International Trading Co., Ltd. | Shanghai | Shanghai | Exporter | 106 |
| Changzhou Biomet Medical Devices Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 54 |
Record
| Listed by | Smith & Nephew, Inc., United States |
|---|---|
| FDA product code | JDS |
| Device type | Nail, Fixation, Bone |
| Device class | Class 2 |
| Regulation | 21 CFR 888.3030 |
| Establishment | Contract manufacturer, Remanufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| 510(k) | K040212 |
| Record published | 2015-08-25 |
| Record revised | 2025-02-06 (version 8) |
Specifications
| MRI safety | MR Conditional |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Label carries | lot or batch number, expiry date |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 71672030 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Autobahn | Globus Medical, Inc. | 1,102 |
| Trigen | Smith & Nephew, Inc. | 1,062 |
| Nail, Fixation, Bone | Synthes Gmbh | 296 |
| Nail, Fixation, Bone | Synthes (U.S.A.) LP | 237 |
| Intertan MAX | Smith & Nephew, Inc. | 235 |
| EXPert Nail | Synthes Gmbh | 225 |
| Trigen MAX | Smith & Nephew, Inc. | 172 |
| Ajax Nail | Auxein medical Private limited | 168 |
| GPC | GPC Medical Limited | 155 |
| 130° Ga-mma Nail | Auxein medical Private limited | 90 |
| Ender Nail | Smith & Nephew, Inc. | 82 |
| Centronail Titanium Humeral Nailing System | Orthofix SRL | 79 |
Showing 12 of 112. All products under Nail, Fixation, Bone
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K243608 | 2025-06-12 |
| K240061 | 2024-09-20 |
| K241804 | 2024-08-22 |
| K230761 | 2023-06-14 |
| K123948 | 2013-03-14 |
| K061783 | 2006-08-10 |
| K061019 | 2006-06-06 |
| K051557 | 2005-06-30 |
| K050938 | 2005-05-04 |
| K040212 | 2004-02-20 |
| K983942 | 1998-12-04 |
| K981529 | 1998-07-09 |
Recalls Naming This Product
6 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Smith & Nephew Orthopaedics AG | 510(k) number | 2022-04-22 | Class II | Under Investigation by firm | Right nails were anodized, marked, and labelled as left nails and vice versa | 17 units | |
| Smith & Nephew, Inc. | 510(k) number | 2020-11-02 | Class II | Packaging process control | Manufacturing packaging error. | 41 units | 2023-01-06 |
| Smith & Nephew, Inc. | 510(k) number | 2017-10-24 | Class II | Employee error | A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down. | 17 units | |
| Smith & Nephew, Inc. | 510(k) number | 2014-07-02 | Class II | Under Investigation by firm | There was an error in manufacturing which resulted in the proximal lag screw hole being positioned incorrectly in several batches of TRIGEN(R); INTERTAN(TM); NAILS. | 601 units | 2019-08-20 |
| Smith & Nephew, Inc. | 510(k) number | 2014-04-04 | Class II | Process control | A manufacturing error resulted in the set screw being inserted too deep in the affected nails. | 1,253 units total | 2017-07-01 |
| Smith & Nephew, Inc. | 510(k) number | 2014-04-04 | Class II | Process control | A manufacturing error resulted in the set screw being inserted too deep in the affected nails. | 1,253 units total | 2017-07-01 |
Recalls
3 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2020-06-24 | Class II | Reprocessing Controls | A manufacturing error which resulted in the re-sterilization of the ultra-high-molecular-weight polyethylene (UHMWPE) sleeve. | 12 units | 2021-04-28 |
| 2013-02-07 | Class II | Labeling mix-ups | 75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label. | 90 units | 2016-01-11 |
| 2013-02-07 | Class II | Labeling mix-ups | 75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label. | 69 units | 2016-01-11 |