TM Reverse Shoulder System
4 catalogue numbers from 82-0180-134-00, listed by Zimmer, Inc., United States, under Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented. 9 manufacturers in China are registered for this product code.
Who Else Makes It
Concentrated in Jiangsu (5), Shanghai (3), Zhejiang (1). 3 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Paragon Medical Device (Changzhou) Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 53 |
| Stryker Orthmed (Changzhou) Medical Instrument Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 10 |
| Stryker Trauson (Changzhou) Medical Technology Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 3 |
| Zimmer (Shanghai) Medical International Trading Co., Ltd. | Shanghai | Shanghai | Exporter | 106 |
| Zhejiang Biomet Medical Products Co., Ltd. | Jinhua | Zhejiang | Contract manufacturer | 74 |
| Shanghai JPY ION-Tech Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 70 |
| Changzhou Biomet Medical Devices Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 54 |
| Arthrex (Shanghai) Co., Ltd. | Shanghai | Shanghai | Exporter | 38 |
| Johnson & Johnson Medical (Depuy – Suzhou) Ltd. | Suzhou | Jiangsu | Manufacturer | 34 |
Record
| Listed by | Zimmer, Inc., United States |
|---|---|
| FDA product code | KWS |
| Device type | Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented |
| Device class | Class 2 |
| Regulation | 21 CFR 888.3660 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Manufacturer, Exporter |
| 510(k) | K052906 |
| Record published | 2017-02-13 |
| Record revised | 2026-02-09 (version 5) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Sterilisation | Moist Heat or Steam Sterilization |
| Prescription | Prescription only |
| Latex | Labelled latex-free |
| Label carries | serial number |
| Sizes | Outer diameter 7.5 mm |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 82-0180-134-00 | In distribution | |
| 82-0180-132-00 | In distribution | |
| 82-0180-135-00 | In distribution | |
| 82-0180-033-03 | Outer Diameter 7.5 mm | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Equinoxe | Exactech, Inc. | 442 |
| Reunion | Stryker GmbH | 321 |
| DJO Surgical | Encore Medical, L.P. | 266 |
| SMR Shoulder | Limacorporate | 209 |
| Tornier Flex Shoulder System | Tornier Inc. | 187 |
| Anatomical Shoulder™ | Zimmer, Inc. | 166 |
| Arthrex® | Arthrex, Inc. | 154 |
| Comprehensive Shoulder System | Biomet Orthopedics | 111 |
| Flex Shoulder System | Tornier Inc. | 109 |
| Aequalis™ PerFORM+ | Tornier Inc. | 108 |
| Aequalis™ Ascend™ Flex | Tornier Inc. | 102 |
| Anatomic Shoulder Prosthesis | Medacta International SA | 90 |
Showing 12 of 326. All products under Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K193180 | 2020-02-24 |
| K191814 | 2019-10-23 |
| K160085 | 2016-03-08 |
| K142403 | 2014-11-24 |
| K062029 | 2006-10-31 |
| K051623 | 2005-07-19 |
| K870262 | 1987-04-03 |
Recalls Naming This Product
12 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Zimmer, Inc. | 510(k) number | 2017-05-22 | Class II | Process control | Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture. | 2825 units | 2019-03-29 |
| Zimmer, Inc. | 510(k) number | 2013-10-17 | Class II | Device Design | As a result of complaints for disassociation of the TM Reverse Glenosphere from the Baseplate, Zimmer updated the related surgical technique and released a video to provide additional instructions to users on how to properly assemble the Gl | 73,002 units | 2014-09-12 |
| Zimmer, Inc. | 510(k) number | 2013-08-01 | Class II | Device Design | Zimmer is initiating a lot specific recall of the Trabecular Metal” Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a result, there is a potenti | 771 | 2015-03-02 |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 9,790 total distal pilots in distribution. | 2012-09-14 |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 9,790 total distal pilots in distribution. | 2012-09-14 |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 13,093 total reamers in distribution. | 2012-09-14 |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 2012-09-14 | |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 2012-09-14 | |
| Zimmer, Inc. | 510(k) number | 2010-09-10 | Not classified | Device Design | The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, dela | 2012-09-14 | |
| Zimmer, Inc. | 510(k) number | 2006-07-20 | Not classified | Other | Components may not lock together as intended. | 197 for all three recalls Z-1416-1418-06 | 2007-12-19 |
| Zimmer, Inc. | 510(k) number | 2006-07-20 | Not classified | Other | Components may not lock together as intended. | 197 for all three recalls Z-1416-1418-06 | 2007-12-19 |
| Zimmer, Inc. | 510(k) number | 2006-07-20 | Not classified | Other | Components may not lock together as intended. | 197 for all three recalls Z-1416-1418-06 | 2007-12-19 |
Recalls
11 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2016-05-17 | Class II | Device Design | In some cases it has been difficult or not possible to disassemble the adjusted Dome Centric from the AS Humeral Rasp after initial positioning of the AS Humeral Trial Head which resulted in the whole construct of the Dome Centric with the | 177 | 2016-10-31 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 45 USA | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Wintetthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 228 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 224 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Wintetthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 139 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 66 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Wintetthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 26 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 164 in USA | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 74 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 260 | 2014-04-16 |
| 2012-09-04 | Class II | Error in labeling | Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient | 54 | 2014-04-16 |
| 2007-06-25 | Not classified | Other | The package actually contains a knee patellar implant. | 10 | 2007-12-05 |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 | |
| 2005-08-23 | Not classified | Other | The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement. | 2005-12-08 |