| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 1125 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 3328 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 540 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 805 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 2372 units | |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 20476 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 41581 | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 24146 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 56468 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 1243 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 32408 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 12774 | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 1757 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 291 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 295 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 585 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 299 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 298 units | 2015-08-05 |
| 2006-02-17 | Not classified | Other | The inserter shafts bend, as they were manufactured with a different stainless steel than specified. | 200 pieces | 2006-06-28 |
| 2006-02-17 | Not classified | Other | The inserter shafts bend, as they were manufactured with a different stainless steel than specified. | 200 pieces | 2006-06-28 |
| 2005-10-27 | Not classified | Other | The device may contain one strand of suture instead of two strands | 19 boxes (5 pieces each) | 2006-05-02 |