| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 1125 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 3328 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 540 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 805 units | |
| 2023-04-13 | Class II | Use error | IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t | 2372 units | |
| 2019-11-04 | Class II | Employee error | Specific product codes and lots of the Mitek HEALIX ADVANCE” Knotless Anchors were manufactured with BIOCRYL” Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective | 299 | 2020-05-05 |
| 2019-11-04 | Class II | Employee error | Specific product codes and lots of the Mitek HEALIX ADVANCE” Knotless Anchors were manufactured with BIOCRYL” Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective | 599 | 2020-05-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 20476 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 41581 | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 24146 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 56468 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 1243 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 32408 units | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 12774 | 2015-08-05 |
| 2012-01-27 | Not classified | Process control | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair | 1757 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 291 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 295 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 585 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 299 units | 2015-08-05 |
| 2011-10-31 | Not classified | Employee error | Incorrect suture configuration | 298 units | 2015-08-05 |
| 2006-02-17 | Not classified | Other | The inserter shafts bend, as they were manufactured with a different stainless steel than specified. | 200 pieces | 2006-06-28 |
| 2006-02-17 | Not classified | Other | The inserter shafts bend, as they were manufactured with a different stainless steel than specified. | 200 pieces | 2006-06-28 |
| 2005-10-27 | Not classified | Other | The device may contain one strand of suture instead of two strands | 19 boxes (5 pieces each) | 2006-05-02 |