Panalok

PANALOK Absorbable Anchor 3.5mm Size 1 (4 metric) PDS II (polydioxanone) suture, 36 Inches (91cm), double-armed with CP-2 needles, with disposable inserter.

Registry build 21.09.26
2manufacturers in China registered for this device type
1catalogue number
3510(k) clearances naming this product code
25recalls on this code

Who Else Makes It

All of them in Shanghai.

Chinese manufacturers registered for FDA product code MAI
Manufacturer City Province Registered as Device types
Zimmer (Shanghai) Medical International Trading Co., Ltd. Shanghai Shanghai Exporter 106
Arthrex (Shanghai) Co., Ltd. Shanghai Shanghai Exporter 38

Record

Regulatory record for Panalok
Listed by DePuy Mitek, Inc., United States
FDA product code MAI
Device type Fastener, Fixation, Biodegradable, Soft Tissue
Device class Class 2
Regulation 21 CFR 888.3030
Establishment Specification developer, Manufacturer
510(k) K952687
Record published 2015-09-01
Record revised 2021-01-28 (version 7)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Supplied sterile Yes
Single use Yes
Prescription Prescription only
Label carries lot or batch number, expiry date
Storage Storage environment temperature up to 80 °F; Handling environment temperature up to 26 °C; Storage environment temperature up to 26 °C; Handling environment temperature up to 80 °F
Distribution end date 2021-01-28

Catalogue Numbers

Catalogue numbers for Panalok
Catalogue number Size Status
210625 Withdrawn

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Arthrex® Arthrex, Inc. 545
Fix2Lock Osteonic Co., Ltd. 118
Delta Stryker Leibinger Gmbh & Co. KG 110
Reunite Screw System Biomet Orthopedics 85
ComposiTCP S.B.M 61
ActivaScrew™ Interference Bioretec Ltd. 54
Ligafix S.B.M 40
Twinfix Smith & Nephew, Inc. 28
Biosure Smith & Nephew, Inc. 27
TorqLoc Interference Bioretec Ltd. 27
FiberFIX Nanova BIO Materials Inc 25
AlphaVent Stryker GmbH 23

Showing 12 of 176. All products under Fastener, Fixation, Biodegradable, Soft Tissue

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K243790 2025-02-05
K241010 2024-09-11
K240441 2024-03-15

Recalls

7 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2023-04-13 Class II Use error IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t 1125 units
2023-04-13 Class II Use error IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t 3328 units
2023-04-13 Class II Use error IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t 540 units
2023-04-13 Class II Use error IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t 805 units
2023-04-13 Class II Use error IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter t 2372 units
2019-11-04 Class II Employee error Specific product codes and lots of the Mitek HEALIX ADVANCE” Knotless Anchors were manufactured with BIOCRYL” Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective 299 2020-05-05
2019-11-04 Class II Employee error Specific product codes and lots of the Mitek HEALIX ADVANCE” Knotless Anchors were manufactured with BIOCRYL” Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective 599 2020-05-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 20476 units 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 41581 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 24146 units 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 56468 units 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 1243 units 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 32408 units 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 12774 2015-08-05
2012-01-27 Not classified Process control Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair 1757 units 2015-08-05
2011-10-31 Not classified Employee error Incorrect suture configuration 291 units 2015-08-05
2011-10-31 Not classified Employee error Incorrect suture configuration 295 units 2015-08-05
2011-10-31 Not classified Employee error Incorrect suture configuration 585 units 2015-08-05
2011-10-31 Not classified Employee error Incorrect suture configuration 299 units 2015-08-05
2011-10-31 Not classified Employee error Incorrect suture configuration 298 units 2015-08-05
2006-02-17 Not classified Other The inserter shafts bend, as they were manufactured with a different stainless steel than specified. 200 pieces 2006-06-28
2006-02-17 Not classified Other The inserter shafts bend, as they were manufactured with a different stainless steel than specified. 200 pieces 2006-06-28
2005-10-27 Not classified Other The device may contain one strand of suture instead of two strands 19 boxes (5 pieces each) 2006-05-02

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker