Itotal® Cruciate Retaining (CR) Knee Replacement System
Catalogue number TCR121111102, listed by ConforMIS, Inc., United States, under Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive. 8 manufacturers in China are registered for this product code.
Who Else Makes It
Concentrated in Jiangsu (4), Shanghai (2), Zhejiang (1). 3 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Zhejiang Biomet Medical Products Co., Ltd. | Jinhua | Zhejiang | Contract manufacturer | 74 |
| Paragon Medical Device (Changzhou) Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 53 |
| b-ONE Medical (Beijing) Co., Ltd. | Beijing | Beijing | Contract manufacturer | 7 |
| Zimmer (Shanghai) Medical International Trading Co., Ltd. | Shanghai | Shanghai | Exporter | 106 |
| Shanghai JPY ION-Tech Co., Ltd. | Shanghai | Shanghai | Contract manufacturer | 70 |
| Changzhou Biomet Medical Devices Co., Ltd. | Changzhou | Jiangsu | Contract manufacturer | 54 |
| Steris Sterilization Technologies (Suzhou) Ltd | Suzhou | Jiangsu | Contract manufacturer | 44 |
| Johnson & Johnson Medical (Depuy – Suzhou) Ltd. | Suzhou | Jiangsu | Manufacturer | 34 |
Record
| Listed by | ConforMIS, Inc., United States |
|---|---|
| FDA product code | OIY |
| Device type | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive |
| Device class | Class 2 |
| Regulation | 21 CFR 888.3560 |
| Establishment | Remanufacturer, Reprocessor, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer |
| 510(k) | K140423 |
| Record published | 2015-10-26 |
| Record revised | 2018-07-06 (version 3) |
Specifications
| MRI safety | MR Conditional |
|---|---|
| Supplied sterile | Yes |
| Single use | Yes |
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | expiry date, serial number |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| TCR121111102 | Patient Specific | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Persona® | Zimmer, Inc. | 552 |
| DJO Surgical | Encore Medical, L.P. | 415 |
| Genflex2 Total Knee System | Globus Medical, Inc. | 389 |
| Persona® Vivacit-E® | Zimmer, Inc. | 332 |
| Truliant, Activit-E | Exactech, Inc. | 272 |
| TriVerse Meniscal Insert PS | Signature Orthopaedics PTY Limited | 220 |
| U2™ Knee | United Orthopedic Corp. | 195 |
| BKS E-Vitalize | Ortho Development Corp. | 193 |
| Vanguard XP Instrumentation | Biomet Orthopedics | 169 |
| Attune | Depuy (Ireland) | 161 |
| Vanguard Complete Knee System | Biomet Orthopedics | 153 |
| EXp Tibial Inserts and Patellar Components | Globus Medical, Inc. | 144 |
Showing 12 of 102. All products under Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive
Recalls Naming This Product
7 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| ConforMIS, Inc. | 510(k) number | 2016-01-28 | Class II | Packaging process control | Incorrect femoral component distributed | 1 unit | 2016-03-09 |
| ConforMIS, Inc. | 510(k) number | 2015-08-31 | Class II | Process design | May contain small amounts of ethylene glycol residue | 68 units | 2016-03-02 |
| ConforMIS, Inc. | 510(k) number | 2015-08-31 | Class II | Process design | May contain small amounts of ethylene glycol residue | 3 units | 2016-03-02 |
| ConforMIS, Inc. | 510(k) number | 2015-08-31 | Class II | Process design | May contain small amounts of ethylene glycol residue | 313 units | 2016-03-02 |
| ConforMIS, Inc. | 510(k) number | 2015-08-31 | Class II | Process design | May contain small amounts of ethylene glycol residue | 348 units | 2016-03-02 |
| ConforMIS, Inc. | 510(k) number | 2015-08-31 | Class II | Process design | May contain small amounts of ethylene glycol residue | 110 units | 2016-03-02 |
| ConforMIS, Inc. | 510(k) number | 2014-11-19 | Class II | Labeling mix-ups | Incorrect serial number of tibial inserts packaged in tray may result in inserts not fitting | 1 unit | 2015-10-08 |
Recalls
1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2022-11-14 | Class II | Process control | Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure. | 70 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.