Vader Pedicle System, K-Wire, Threaded, Nitinol, Ø 1.6 X 600 Mm
Catalogue number 42-546-1, listed by Icotec AG, Switzerland, under Bone Cement, Posterior Screw Augmentation. 2 manufacturers in China are registered for this product code.
Who Else Makes It
Concentrated in Jiangsu (1), Shandong (1). One is registered as a contract manufacturer, able to build to a specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Shandong Weigao Orthopaedic Device Co., Ltd. | Weihai | Shandong | Contract manufacturer | 42 |
| Medtronic (Changzhou) Medical Devices Technology Co., Ltd. | Changzhou | Jiangsu | Manufacturer | 53 |
Record
| Listed by | Icotec AG, Switzerland |
|---|---|
| FDA product code | PML |
| Device type | Bone Cement, Posterior Screw Augmentation |
| Device class | Class 2 |
| Regulation | 21 CFR 888.3027 |
| Establishment | Complaint-file establishment, Manufacturer |
| Record published | 2024-11-22 |
| Record revised | 2026-03-19 (version 3) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Sterilisation | Moist Heat or Steam Sterilization |
| Prescription | Prescription only |
| Label carries | lot or batch number, manufacturing date |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 42-546-1 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| ReLine | Nuvasive, Inc. | 321 |
| M.u.s.t. Pedicle Screw System | Medacta International SA | 248 |
| CarboClear Hybrid Pedicle Screw System | CarboFix Orthopedics, Ltd. | 153 |
| Viper | Medos International Sàrl | 32 |
| Expedium Verse | Medos International Sàrl | 26 |
| CarboClear X Pedicle Screw System | CarboFix Orthopedics, Ltd. | 18 |
| Vader Pedicle System, Screw, Carbon/PEEK, Polyax., Fen., Ø 5.5 x 55 mm | Icotec AG | 13 |
| CarboClear Modular Hybrid Pedicle Screw System | CarboFix Orthopedics, Ltd. | 11 |
| CarboClear Pedicle Screw System | CarboFix Orthopedics, Ltd. | 9 |
| icotec Pedicle System, Screw, Ti, Polyax., Fen., Ø 8.5 x 45 mm | Icotec AG | 6 |
| Vader Pedicle System, K-wire Dispenser, Extra Long, Ø 1.6 x 600 mm, Ø 1.1 x 600 | Icotec AG | 1 |
| Romeo® 2 | Spineart SA | 1 |
Showing 12 of 14. All products under Bone Cement, Posterior Screw Augmentation
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K222789 | 2023-01-09 |
| K200596 | 2020-10-13 |
Recalls
2 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2023-11-28 | Class II | Other | icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighte | |
| 2023-11-28 | Class II | Other | icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighte | 79 devices (US only) |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.