Artis Q Floor

Catalogue number 10848280, listed by Siemens Healthcare Limited, Canada, under System, X-Ray, Angiographic. 7 manufacturers in China are registered for this product code.

Registry build 21.09.26
7manufacturers in China registered for this device type
1catalogue number
5510(k) clearances naming this product code

Who Else Makes It

Concentrated in Shanghai (2), Beijing (2), Guangdong (2).

Chinese manufacturers registered for FDA product code IZI
Manufacturer City Province Registered as Device types
Arthrex (Shanghai) Co., Ltd. Shanghai Shanghai Exporter 38
GE Hangwei Medical Systems Co., Ltd. Beijing Beijing Specification developer 22
GE Healthcare (China) Co., Ltd. Beijing Beijing Specification developer 22
Scivita Medical Technology Co., Ltd. Suzhou Jiangsu Manufacturer 19
Siemens Shanghai, Medical Equipment Ltd. Shanghai Shanghai Contract manufacturer 10
Guangdong Optomedic Technologies, Inc. Foshan Guangdong Manufacturer 7
Zhuhai Dipu Medical Technology Co., Ltd. Zhuhai Guangdong Manufacturer 3

Record

Regulatory record for Artis Q floor
Listed by Siemens Healthcare Limited, Canada
FDA product code IZI
Device type System, X-Ray, Angiographic
Device class Class 2
Regulation 21 CFR 892.1600
Establishment Contract manufacturer, Manufacturer, Exporter
510(k) K123529
Record published 2016-09-24
Record revised 2021-11-17 (version 7)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number
Storage Storage environment temperature -20 to 70 °C; Handling environment humidity 20 to 75 % RH; Storage environment humidity 10 to 95 % RH; Storage environment atmospheric pressure 70 to 106 KiloPascal; Handling environment temperature 15 to 30 °C; Handling environment atmospheric pressure 70 to 106 KiloPascal

Catalogue Numbers

Catalogue numbers for Artis Q floor
Catalogue number Size Status
10848280 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Sterimed Sterimed, Inc. 114
Med-Syringe Argon Medical Devices, Inc. 16
Control Syringe Argon Medical Devices, Inc. 14
Novadaq and Pinpoint Novadaq Technologies, Inc. 12
DeRoyal Deroyal Industries, Inc. 12
Namic Navilyst Medical, Inc. 9
Namic Medline Industries, Inc. 8
iMRX System Imris Imaging, Inc. 6
System, X-Ray, Angiographic B. Braun Medical Inc. 4
System, X-Ray, Angiographic Intuitive Surgical, Inc. 3
Leica FL800 Leica Microsystems (Schweiz) AG 3
Infrared 800 with Flow® 800 Option Carl Zeiss Meditec Inc 2

Showing 12 of 51. All products under System, X-Ray, Angiographic

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K010721 2001-03-30
K993947 1999-12-21
K812721 1981-11-04
K811149 1981-06-09
K781991 1978-12-07

Recalls Naming This Product

90 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Artis Q floor
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Siemens Medical Solutions USA, Inc. 510(k) number 2024-03-21 Class II Device Design Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break. the system cannot be 583 units
Siemens Medical Solutions USA, Inc. 510(k) number 2022-12-16 Class II Software design In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternati 609 US; 1627 worldwide
Siemens Medical Solutions USA, Inc. 510(k) number 2022-08-10 Class II Device Design Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to rele 1,389 (590 US; 799 OUS) (Total)
Siemens Medical Solutions USA, Inc. 510(k) number 2022-05-24 Class II Component design/selection It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only b 233 worldwide; 32 US 2025-01-22
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 305 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 709 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 277 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 373 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 24 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 7 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 183 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 80 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 418 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 269 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 38 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 33 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 81 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2022-02-23 Class II Device Design If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message “TUBE HOT, have a break”. Several minut 69 systems
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 3 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 8 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 19 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 2 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 8 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 67 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 15 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 3 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 15 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-10-20 Class II Software design If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure 2 units 2024-06-18
Siemens Medical Solutions USA, Inc. 510(k) number 2021-09-24 Class II Software design Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5 de 43 US
Siemens Medical Solutions USA, Inc. 510(k) number 2021-06-30 Class II Software design Due to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart,
and result in delay of clinical treatment
70 US (172 WW) 2023-02-10
Siemens Medical Solutions USA, Inc. 510(k) number 2021-03-30 Class II Equipment maintenance Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited function 3,320 units U.S. 2022-07-12
Siemens Medical Solutions USA, Inc. 510(k) number 2021-03-26 Class II Device Design Some Artis zee ceiling systems show an increased abrasion of the cabling at the cable outlet of the inner C-arm, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where it 772 units U.S.
Siemens Medical Solutions USA, Inc. 510(k) number 2020-11-30 Class II Software design When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro “Fluoro Override”, the measured Air Kerma Rate can exceed the regulatory limits. 5602
Siemens Medical Solutions USA, Inc. 510(k) number 2020-10-21 Class II Employee error Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation 26 distributed worldwide (4 in US) 2021-05-05
Siemens Medical Solutions USA, Inc. 510(k) number 2020-08-03 Class II Software Manufacturing/Software Deployment Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause co 124 U.S 2021-02-03
Siemens Medical Solutions USA, Inc. 510(k) number 2020-04-30 Class II Nonconforming Material/Component Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated an 15 US 2020-09-11
Siemens Medical Solutions USA, Inc. 510(k) number 2020-02-14 Class II Software design A software issue could potentially cause the stand and table movements to be blocked. 1452 worldwide (638 US) 2021-04-20
Siemens Medical Solutions USA, Inc. 510(k) number 2020-02-14 Class II Software design A software issue could potentially cause the stand and table movements to be blocked. 1452 worldwide (638 US) 2021-04-20
Siemens Medical Solutions USA, Inc. 510(k) number 2020-02-14 Class II Software design A software issue could potentially cause the stand and table movements to be blocked. 1452 worldwide (638 US) 2021-04-20
Siemens Medical Solutions USA, Inc. 510(k) number 2019-11-14 Class II Software design In affected Artis zee floor Interventional Fluoroscopic X-Ray Systems, the activation of a collision sensor will cause a block the movement of the floating tabletop and could result in a delay or interruption of a procedure. 3 2020-05-12

Showing 40 of 90.

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.