Innova
Innova IGS 620 002 UDI
Who Else Makes It
Concentrated in Shanghai (2), Beijing (2), Guangdong (2).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Arthrex (Shanghai) Co., Ltd. | Shanghai | Shanghai | Exporter | 38 |
| GE Hangwei Medical Systems Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| GE Healthcare (China) Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| Scivita Medical Technology Co., Ltd. | Suzhou | Jiangsu | Manufacturer | 19 |
| Siemens Shanghai, Medical Equipment Ltd. | Shanghai | Shanghai | Contract manufacturer | 10 |
| Guangdong Optomedic Technologies, Inc. | Foshan | Guangdong | Manufacturer | 7 |
| Zhuhai Dipu Medical Technology Co., Ltd. | Zhuhai | Guangdong | Manufacturer | 3 |
Record
| Listed by | GE Medical Systems, LLC, United States |
|---|---|
| FDA product code | IZI |
| Device type | System, X-Ray, Angiographic |
| Device class | Class 2 |
| Regulation | 21 CFR 892.1600 |
| Establishment | Specification developer, Manufacturer |
| 510(k) | K122457 |
| Record published | 2016-09-24 |
| Record revised | 2025-08-20 (version 5) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | manufacturing date, serial number |
| Distribution end date | 2024-05-01 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| UPG X100 001 | Withdrawn | |
| IGS 620 002 | Withdrawn |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Sterimed | Sterimed, Inc. | 114 |
| Med-Syringe | Argon Medical Devices, Inc. | 16 |
| Control Syringe | Argon Medical Devices, Inc. | 14 |
| Novadaq and Pinpoint | Novadaq Technologies, Inc. | 12 |
| DeRoyal | Deroyal Industries, Inc. | 12 |
| Namic | Navilyst Medical, Inc. | 9 |
| Namic | Medline Industries, Inc. | 8 |
| iMRX System | Imris Imaging, Inc. | 6 |
| System, X-Ray, Angiographic | B. Braun Medical Inc. | 4 |
| System, X-Ray, Angiographic | Intuitive Surgical, Inc. | 3 |
| Leica FL800 | Leica Microsystems (Schweiz) AG | 3 |
| Infrared 800 with Flow® 800 Option | Carl Zeiss Meditec Inc | 2 |
Showing 12 of 51. All products under System, X-Ray, Angiographic
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K974367 | 1998-02-13 |
| K973270 | 1997-11-25 |
| K973368 | 1997-10-17 |
| K960575 | 1996-05-07 |
| K945375 | 1995-01-30 |
| K943355 | 1994-11-07 |
| K942120 | 1994-07-18 |
| K941946 | 1994-05-20 |
| K926258 | 1993-03-15 |
| K905653 | 1991-05-15 |
Recalls Naming This Product
20 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| GE Medical Systems, LLC | 510(k) number | 2019-03-15 | Class II | Device Design | There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously. | 49 devices | |
| GE Medical Systems SCS | 510(k) number | 2018-03-29 | Class II | Process control | The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use. | 102 devices total | 2021-06-15 |
| GE Medical Systems SCS | 510(k) number | 2018-03-29 | Class II | Process control | The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use. | 102 devices total | 2021-06-15 |
| GE Medical Systems, LLC | 510(k) number | 2015-12-18 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor. | 2,111 (1078 US; 1033 OUS) | 2016-09-07 |
| GE Medical Systems, LLC | 510(k) number | 2015-11-13 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems. | 108 | 2016-09-07 |
| GE Medical Systems, LLC | 510(k) number | 2015-11-13 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue related to a non-recoverable loss of displayed imaging (loss of monitor video) involving Optima CL323i & Optima IGS 320 systems. | 45 | 2016-09-07 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 3 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 4 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 13 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 1 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 6 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 2 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 10 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 17 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 6 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 3 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 8 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2015-05-29 | Class II | Nonconforming Material/Component | GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery). | 7 | 2015-10-28 |
| GE Medical Systems, LLC | 510(k) number | 2014-09-04 | Class II | Device Design | Loss of imaging – affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism. | 10 (4 US, 6 OUS). | 2015-01-21 |
| GE Medical Systems, LLC | 510(k) number | 2013-11-15 | Class II | Software design | GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the us | 28 (12 US, 16 OUS). | 2014-03-28 |
Recalls
9 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2019-03-15 | Class II | Device Design | There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously. | 1 device | |
| 2019-03-15 | Class II | Device Design | There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously. | 22 devices | |
| 2016-11-21 | Class II | Nonconforming Material/Component | GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors – Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems. | Total for all products 1,598 (USA 824, OUS 774) | 2017-09-26 |
| 2016-11-21 | Class II | Nonconforming Material/Component | GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors – Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems. | Total for all products 1,598 (USA 824, OUS 774) | 2017-09-26 |
| 2016-11-21 | Class II | Nonconforming Material/Component | GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors – Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems. | Total for all products 1,598 (USA 824, OUS 774) | 2017-09-26 |
| 2016-11-21 | Class II | Nonconforming Material/Component | GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors – Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems. | Total for all products 1,598 (USA 824, OUS 774) | 2017-09-26 |
| 2011-12-20 | Class II | Radiation Control for Health and Safety Act | It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems. | 199 installed in the US | 2013-09-18 |
| 2011-12-20 | Class II | Radiation Control for Health and Safety Act | It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems. | 199 installed in the US | 2013-09-18 |
| 2011-12-20 | Class II | Radiation Control for Health and Safety Act | It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems. | 199 installed in the US | 2013-09-18 |
| 2011-05-04 | Not classified | Radiation Control for Health and Safety Act | It was discovered that when the Diamentor M4-KDK is installed on an Innova system the dose display is disabled on the certified Innova monitor and enabled on the Diamentor M4-KDK which is not certified as required by FDA regulation (Subpar | 19 | 2012-03-29 |
| 2009-04-16 | Not classified | Device Design | Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monit | 60 | 2015-12-03 |
| 2008-01-14 | Not classified | Process control | Risk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system | 252 | 2011-12-17 |
| 2004-03-24 | Not classified | Other | Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox. | 2006-07-24 | |
| 2004-03-24 | Not classified | Other | Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox. | 2006-07-24 | |
| 2004-03-24 | Not classified | Other | Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox. | 2006-07-24 | |
| 2004-03-24 | Not classified | Other | Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox. | 2006-07-24 | |
| 2004-03-24 | Not classified | Other | Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox. | 2006-07-24 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.