Regulatory record for Philips Medical Systems DMC
Listed by Villa Sistemi Medicali SPA
FDA product code KPW
Device type Device, Beam Limiting, X-Ray, Diagnostic
Device class Class 2
Regulation 21 CFR 892.1610
510(k) K050190
Record published 2016-09-24
Record revised 2019-08-05 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number
Distribution end date 2019-07-31

Catalogue Numbers

Catalogue numbers for Philips Medical Systems DMC
Catalogue number Size Status
Juno DRF Withdrawn

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Ralco srl Ralco SRL 10
Villa Sistemi Medicali Villa Sistemi Medicali SPA 2
X-ray beam limiting device Canon Inc. 1
Device, Beam Limiting, X-Ray, Diagnostic Yoshida Denzai Kogyo Co., Ltd. 1
Collimator AL02 II eL Siemens Healthcare Limited 1
X-ray Beam Filter Kit Carestream Health, Inc. 1
Collimator ML01 II Siemens Healthcare Limited 1

Recalls Naming This Product

14 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Philips Medical Systems DMC
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Villa Sistemi Medicali S.P.A. 510(k) number 2019-11-04 Class II Device Design An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditi 46
Villa Sistemi Medicali S.P.A. 510(k) number 2019-11-04 Class II Device Design An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditi 3
Villa Sistemi Medicali S.P.A. 510(k) number 2019-11-04 Class II Device Design An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditi 7
Villa Sistemi Medicali S.P.A. 510(k) number 2019-11-04 Class II Device Design An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditi 1
Villa Sistemi Medicali S.P.A. 510(k) number 2016-12-27 Class II Nonconforming Material/Component Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position. 65 devices in total 2021-03-09
Villa Sistemi Medicali S.P.A. 510(k) number 2016-12-27 Class II Nonconforming Material/Component Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position. 65 devices in total 2021-03-09
Villa Sistemi Medicali S.P.A. 510(k) number 2016-12-27 Class II Nonconforming Material/Component Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position. 65 devices in total 2021-03-09
Villa Sistemi Medicali S.P.A. 510(k) number 2016-12-27 Class II Nonconforming Material/Component Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position. 65 devices in total 2021-03-09
Acteon, Inc. 510(k) number 2016-02-19 Class II Device Design The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) T
66 distributed in US
Acteon, Inc. 510(k) number 2016-02-19 Class II Device Design The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) T
66 distributed in US
Acteon, Inc. 510(k) number 2016-02-19 Class II Device Design The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) T
66 distributed in US
Acteon, Inc. 510(k) number 2016-02-19 Class II Device Design The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) T
66 distributed in US
Acteon, Inc. 510(k) number 2014-12-14 Class II Radiation Control for Health and Safety Act The system does not provide the appropriate audible signal, permanent activation, and manual override, although the system is in high-level control functionality 30 units distributed in the US 2017-03-16
Acteon, Inc. 510(k) number 2013-08-02 Class II Radiation Control for Health and Safety Act It has been discovered that the Air Kerma rate information is inconsistent in Continuous Fluoro modes 30×30 (12″x12″) and 15×15 (6″x6). This could result in a deviation of the displayed values of Air Kerma Rate and Cumulated Air Kerma from 22 units within US 2016-07-25

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer