HAD1717MC
HAD1717MC is a digital flat-panel X-ray detector for general projection radiographic applications. The filing states that it is intended for applications where conventional screen-film or…
HAD1717MC is a digital flat-panel X-ray detector for general projection radiographic applications. The filing states that it is intended for applications where conventional screen-film or computed-radiography systems may be used, and explicitly excludes mammographic applications.
The detector digitizes signals received from its imaging matrix and transmits the digital signal to a PC. Its software can acquire and store digital images and display them for viewing, giving the detector a defined image-capture and transfer role.
Listed Item
The HAD1717MC is a wired digital flat panel X-ray detector. The HAD1717MC X-ray detector consists of a TFT panel and a scintillator electronic controller, and supports AED (Automatic Exposure Detection) function that can operate without a separate trigger signal. Flat panel sensors for HAD1717MC are fabricated using thin film technology based on amorphous silicon technology, and the sensors are similar to existing photodiode arrays. In this case, each pixel is composed of a photodiode and a thin film transistor (TFT). Amorphous silicon photodiodes are optimized for green wavelengths and can be used in combination with scintillators such as CsI. The incident X-rays are converted into visible light by the scintillator, and the visible light is stored in an internal capacitor through a photodiode. Charges charged to capacitors switch TFT gates of panel matrix (panel matrix) and convert them into digital signals on a line and pass them to the control unit. The digital data generated is then passed to the PC via Ethernet.
Image Capture and Viewing
The workstation software receives the detector’s digitized signal on a PC and can acquire, store and display images for viewing. The indication is for general projection radiography and excludes mammographic applications.
Who Else Makes It
Concentrated in Shanghai (5), Jiangsu (4), Beijing (2). 3 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Shantou Institute OF Ultrasonic Instruments Co., Ltd. | Shantou | Guangdong | Contract manufacturer | 8 |
| iRay Imaging Technology (Haining) Limited | Haining | Zhejiang | Contract manufacturer | 6 |
| iRay Imaging Technology Taicang Limited | Taicang | Jiangsu | Contract manufacturer | 3 |
| GE Hangwei Medical Systems Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| GE Healthcare (China) Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| Shanghai United Imaging Healthcare Co., Ltd. | Shanghai | Shanghai | Manufacturer | 12 |
| Siemens Shanghai, Medical Equipment Ltd. | Shanghai | Shanghai | Contract manufacturer | 10 |
| Carestream Health, Inc., Global R&D Center | Shanghai | Shanghai | Specification developer | 8 |
| Philips Healthcare (Suzhou) Co., Ltd. | Suzhou | Jiangsu | Contract manufacturer | 5 |
| iRay Group | Shanghai | Shanghai | Specification developer | 4 |
| Fujifilm Imaging System (Suzhou) Co., Ltd. | Suzhou | Jiangsu | Manufacturer | 3 |
| Careray Digital Medical Technology Co., Ltd. | Suzhou | Jiangsu | Manufacturer | 2 |
Showing 12 of 13. All 13 manufacturers in China registered for Solid State X-Ray Imager (Flat Panel/Digital Imager)
Record
| Listed by | H&Abyz, South Korea |
|---|---|
| FDA product code | MQB |
| Device type | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Device class | Class 2 |
| Regulation | 21 CFR 892.1680 |
| Establishment | Contract manufacturer, Manufacturer |
| 510(k) | K200018 |
| Record published | 2023-04-07 |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | lot or batch number, manufacturing date, serial number |
| Storage | Handling environment temperature 5 to 40 °C; Handling environment humidity 30 to 75 % RH; Handling environment atmospheric pressure 70 to 106 KiloPascal; Storage environment temperature -10 to 50 °C; Storage environment humidity 10 to 80 % RH; Storage environment atmospheric pressure 70 to 106 KiloPascal |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| HAD1717MC | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Sedecal | Sedecal., Sa. | 70 |
| Schick33 | Sirona Dental, Inc. | 44 |
| Drgem | Drgem Corporation | 41 |
| Schick | Dentsply LLC | 35 |
| Wireless Digital Flat Panel Detector | iRay Group | 27 |
| Fuji IP Cassette CC | Fujifilm Corporation | 20 |
| Canon Inc. | Canon Inc. | 20 |
| Fona Stellaris | Sirona Dental, Inc. | 20 |
| Exprimer | Drtech Corporation | 18 |
| Solid State X-Ray Imager (Flat Panel/Digital Imager) | Samsung Electronics Co., Ltd. | 17 |
| X-ray Flat Panel Detector | Careray Digital Medical Technology Co., Ltd. | 16 |
| Regius Cassette | Konica Minolta, Inc. | 16 |
Showing 12 of 509. All products under Solid State X-Ray Imager (Flat Panel/Digital Imager)
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K223930 | 2023-03-06 |
| K213497 | 2021-11-15 |
| K203188 | 2021-03-12 |
| K200018 | 2020-01-31 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.