Azurion
Biplane Interventional X-Ray System with floor-mounted frontal stand (20-inch flat detector), ceiling-mounted lateral stand (12-inch or 15-inch flat detector) and optional interface for an…
Biplane Interventional X-Ray System with floor-mounted frontal stand (20-inch flat detector), ceiling-mounted lateral stand (12-inch or 15-inch flat detector) and optional interface for an OR table.
Who Else Makes It
Concentrated in Beijing (2), Shanghai (2), Guangdong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| GE Hangwei Medical Systems Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| GE Healthcare (China) Co., Ltd. | Beijing | Beijing | Specification developer | 22 |
| Shanghai United Imaging Healthcare Co., Ltd. | Shanghai | Shanghai | Manufacturer | 12 |
| Siemens Shanghai, Medical Equipment Ltd. | Shanghai | Shanghai | Contract manufacturer | 10 |
| Siemens Shenzhen Magnetic Resonance Ltd. | Shenzhen | Guangdong | Contract manufacturer | 8 |
Record
| Listed by | Philips Medical Systems Nederland B.V., Netherlands |
|---|---|
| FDA product code | OWB |
| Device type | Interventional Fluoroscopic X-Ray System |
| Device class | Class 2 |
| Regulation | 21 CFR 892.1650 |
| Establishment | Contract manufacturer, Remanufacturer, Specification developer, Complaint-file establishment, Manufacturer, Exporter |
| Record published | 2017-03-31 |
| Record revised | 2025-12-15 (version 8) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | serial number |
| Storage | Storage environment temperature -13 to 158 °F; Storage environment temperature -25 to 70 °C; Storage environment humidity 5 to 95 % RH; Storage environment atmospheric pressure 70 to 110 KiloPascal |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| Azurion 3 M15 | In distribution | |
| Azurion 7 B20 | In distribution | |
| Azurion 7 B12 | In distribution | |
| Azurion 7 M12 | In distribution | |
| Azurion 5 M12 | In distribution | |
| Azurion 3 M12 | In distribution | |
| Azurion 5 M12 | In distribution | |
| Azurion 7 B20 | In distribution | |
| Azurion 7 M20 | In distribution | |
| Azurion 7 B12 | In distribution | |
| Azurion 7 M12 | In distribution | |
| Azurion 5 M20 | In distribution | |
| Azurion 7 B20 | In distribution | |
| Azurion 7 B12 | In distribution | |
| Azurion 3 M12 | In distribution | |
| Azurion 7 M12 | In distribution | |
| Azurion 3 M15 | In distribution | |
| Azurion 7 M20 | In distribution | |
| Azurion 7 M20 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| TrackX | Trackx Technology Inc. | 86 |
| Trinias | Shimadzu Corp. | 54 |
| Allura Xper FD | Philips Medical Systems Nederland B.V. | 21 |
| LessRay | Nuvasive, Inc. | 16 |
| Intelli-C Accessory | Nrt – Nordisk Røntgen Teknik A/S | 13 |
| Pulse | Nuvasive, Inc. | 12 |
| O-ARM® | Medtronic Navigation, Inc. | 11 |
| Genoray | Genoray Co., Ltd. | 10 |
| Fujifilm Healthcare Americas Corporation | A.T.S. Applicazione Tecnologie Speciali SRL | 7 |
| Fujifilm Medical Systems U.S.A. Inc. | A.T.S. Applicazione Tecnologie Speciali SRL | 6 |
| Stephanix | A.T.S. Applicazione Tecnologie Speciali SRL | 6 |
| Amrad Medical | A.T.S. Applicazione Tecnologie Speciali SRL | 6 |
Showing 12 of 122. All products under Interventional Fluoroscopic X-Ray System
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K254205 | 2026-08-07 |
| K232420 | 2024-02-16 |
| K212813 | 2021-10-01 |
| K201583 | 2021-02-09 |
| K200917 | 2020-05-01 |
| K183101 | 2018-12-31 |
| K183040 | 2018-12-06 |
| K181830 | 2018-08-02 |
| K172822 | 2017-11-22 |
| K172307 | 2017-10-24 |
| K170130 | 2017-06-02 |
| K163715 | 2017-01-26 |
| K162859 | 2016-12-02 |
| K160455 | 2016-08-25 |
| K160147 | 2016-08-10 |
| K161563 | 2016-07-29 |
| K161839 | 2016-07-29 |
| K151598 | 2015-08-17 |
| K142273 | 2015-01-30 |
| K142126 | 2014-12-09 |
| K133819 | 2014-07-10 |
| K140138 | 2014-06-10 |
| K133292 | 2014-03-05 |
| K132147 | 2013-12-16 |
| K132305 | 2013-10-30 |
| K130893 | 2013-09-06 |
| K131263 | 2013-07-24 |
| K130638 | 2013-06-28 |
| K121772 | 2013-03-21 |
Recalls Naming This Product
59 recalls name this product by its own identifiers, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|---|---|
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-07-16 | Class II | Software design | Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements | 10872 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-07-16 | Class II | Software design | Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geo | 2016 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-06-17 | Class II | Under Investigation by firm | Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. | 3,341 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-04-30 | Class II | Under Investigation by firm | It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the aff | 3069 units (872 US, 2224 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-04-28 | Class II | Device Design | Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module. | 5,537 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-04-28 | Class II | Component design/selection | The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. Whe | 1651 units (592 US, 1359 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-31 | Class II | Device Design | Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated. | 1,718 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-19 | Class II | Labeling design | Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion f | 1,033 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 298 units (27 US, 271 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 798 units (42 US, 756 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 670 units (106 US, 564 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 1523 units (322 US, 1201 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 1994 units (360 US, 1634 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 5167 units (1728 US, 3439 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 345 units (10 US, 335 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 920 units (168 US, 752 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-12-12 | Class II | Software design | Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment o | 10,466 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-12-12 | Class II | Software design | Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 – System remains | 228 units (all OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-12-08 | Class II | Software design | Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition | 990 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-06-04 | Class II | Device Design | Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose. | 78 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-05-28 | Class II | Software design | A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identifie | 9,707 units (2300 US, 7407 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 9 units US; 101 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 42 units US; 661 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 6 units US; 246 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 68 units US; 438 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 90 units US; 458 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 251 units US; 938 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 324 units US; 1390 units OUS |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class II | Device Design | Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longit | 1346 units US; 2582 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 263 units (26 US, 237 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 735 units (42 US, 693 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 282 units (9 US, 273 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 632 units (77 US, 555 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 602 units (94 US, 508 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 1,314 units (270 US, 1044 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 1,837 units (337 US, 1500 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2025-01-13 | Class I | Under Investigation by firm | Possibility of the patient falling from the table related to the mattress used on the patient table. | 4,221 units (1480 US, 2741 OUS) |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2024-08-06 | Class II | Under Investigation by firm | The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, becaus | 927 systems |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2024-07-15 | Class II | Device Design | Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart | 3011 units |
| Philips Medical Systems Nederland B.V. | UDI-DI | 2024-02-06 | Class II | Under Investigation by firm | Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to powe | 14 units |
Showing 40 of 59.
Recalls
25 of these are Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2026-06-29 | Class II | Software design | Firmware issue potentially leading to loss of geometry movements. | 4,252 units |
| 2026-06-17 | Class II | Under Investigation by firm | Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. | 5,012 units |
| 2026-06-17 | Class II | Under Investigation by firm | Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. | 257 units |
| 2026-05-18 | Class II | Radiation Control for Health and Safety Act | Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection | |
| 2026-04-30 | Class II | Under Investigation by firm | It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the aff | |
| 2026-04-28 | Class II | Component design/selection | The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. Whe | 3314 units (816 US, 2498 OUS) |
| 2026-04-28 | Class II | Component design/selection | The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. Whe | 25 units (11 US, 14 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 296 units (68 US, 228 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 5 units (OUS only) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 461 units (198 US, 263 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 111 units (31 US, 80 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 5084 units (1616 US, 3468 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 85 units (20 US, 65 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 2 units (2 US, 6 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 2291 units (493 US, 1798 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 10 units (8 US, 2 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 196 units (73 US, 223 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 4 units (2 US, 2 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 2 units (1 US, 1 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 15 units (7 US, 8 OUS) |
| 2026-03-03 | Class II | Device Design | Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. | 112 units (138 US, 422 OUS) |
| 2026-01-26 | Class II | Device Design | Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. | 291 units (27 US, 264 OUS) |
| 2026-01-26 | Class II | Device Design | Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. | 802 units (45 US, 757 OUS) |
| 2026-01-26 | Class II | Device Design | Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. | 655 units (106 US, 549 OUS) |
| 2026-01-26 | Class II | Device Design | Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed – even when a table lock is active. | 1503 units (326 US, 1177 OUS) |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.