| Maquet Critical Care AB | 510(k) number | 2026-06-24 | Class II | Under Investigation by firm | Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness. | 408 US; 11,342 OUS | |
| Getinge Usa Sales Inc | 510(k) number | 2025-01-15 | Class I | Device Design | Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in | 6 units | |
| Getinge Usa Sales Inc | 510(k) number | 2024-05-01 | Class I | Under Investigation by firm | Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which i | 120 units (US) | |
| Getinge Usa Sales Inc | 510(k) number | 2022-12-22 | Class I | Software Design Change | Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia. | 8,016 Total (7,766 OUS; 250 US) | |
| Getinge Usa Sales Inc | 510(k) number | 2022-12-22 | Class I | Software Design Change | Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia. | 8.016 Total (7.766 OUS; 250 US) | |
| Getinge Usa Sales Inc | 510(k) number | 2022-12-22 | Class I | Software Design Change | Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia. | 8.016 Total (7.766 OUS; 250 US) | |
| Getinge Usa Sales Inc | 510(k) number | 2022-11-07 | Class II | Component design/selection | The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could req | 65 US | |
| Getinge Usa Sales Inc | 510(k) number | 2022-05-02 | Class I | Component design/selection | Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated. | 17 units US | 2024-08-06 |
| Getinge Usa Sales Inc | 510(k) number | 2022-05-02 | Class I | Component design/selection | Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated. | 47 units US | 2024-08-06 |
| Getinge Usa Sales Inc | 510(k) number | 2021-12-09 | Class I | Under Investigation by firm | A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride. | 50 US; 898 OUS | |
| Getinge Group Logistics America, LLC | 510(k) number | 2021-01-29 | Class II | Device Design | Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine. | 144 units | 2023-09-28 |