SunMed LLC

AirFlow Adult Resuscitator with B/V Filter, Bag Reservoir, Mask, Attached PEEP Valve and 14' Tubing

Registry build 21.09.26
8manufacturers in China registered for this device type
1catalogue number
40recalls on this code

Who Else Makes It

Concentrated in Fujian (4), Zhejiang (1), Guangdong (1). 5 of them are registered as contract manufacturers, able to build to a buyer’s own specification.

Chinese manufacturers registered for FDA product code BTM
Manufacturer City Province Registered as Device types
Flexicare Medical (Dongguan) Ltd. Dongguan Guangdong Contract manufacturer 48
GaleMed (Xiamen) Co., Ltd. Xiamen Fujian Contract manufacturer 33
Polymed (Xiamen) Plastic Industrial Co., Ltd. Xiamen Fujian Contract manufacturer 17
VR Medical Technology Co., Ltd. Kunshan Jiangsu Contract manufacturer 13
Ambu Limited Xiang Yu F.T.Z., Xiamen Fujian Contract manufacturer 11
Huzhou Junda Industry Co., Ltd. Anji Zhejiang Contract manufacturer 129
Laerdal Medical (Suzhou) Co., Ltd. 215009 Suzhou Jiangsu Province Repacker / relabeller 12
Xiamen Compower Medical Tech. Co., Ltd. Xiamen Fujian Contract manufacturer 9

Record

Regulatory record for SunMed LLC
Listed by SunMed Group Holdings, LLC, United States
FDA product code BTM
Device type Ventilator, Emergency, Manual (Resuscitator)
Device class Class 2
Regulation 21 CFR 868.5915
Establishment Contract manufacturer, Reprocessor, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer
Record published 2023-10-24

Specifications

Specifications
MRI safety MR Conditional
Single use Yes
Prescription Prescription only
Latex Labelled latex-free
Label carries lot or batch number, expiry date, manufacturing date
Packaging Case of 10

Catalogue Numbers

Catalogue numbers for SunMed LLC
Catalogue number Size Status
AF5040MB-PWT In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Ventlab Ventlab, LLC 405
CPR-2 Bag Mercury Medical Enterprises, LLC 231
CPR Bag Mercury Medical Enterprises, LLC 187
Vital Signs™ Airlife, Inc. 102
Medline Medline Industries, Inc. 78
Spur® II Pediatric ResuscitatorSingle Patient Use Resuscitator Ambu A/S 74
Spur® II Adult ResuscitatorSingle Patient Use Resuscitator Ambu A/S 66
Hudson RCI Teleflex Medical LLC 56
Medline Industries, Inc. Medline Industries, Inc. 53
Portex ICU Medical, Inc 45
Spur® II Infant ResuscitatorSingle Patient Use Resuscitator Ambu A/S 45
Curaplex® Bound Tree Medical, LLC 38

Showing 12 of 342. All products under Ventilator, Emergency, Manual (Resuscitator)

Recalls

14 of these are Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2025-05-01 Class I Process control Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which 11,358
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2023-10-11 Class II Device Design A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 350 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 110 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 361 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 80 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 55 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 60 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 40 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 113 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 76 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 1,742 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 60 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 30 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 5 cases 2020-05-12
2018-05-24 Class II Mixed-up of materials/components There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient. 2 cases 2020-05-12
2014-05-16 Class I Process change control The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim. 251,384 2015-11-17
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2012-07-11 Class I Component design/selection Possible volume leakage through the inlet valve during compressions of the resuscitator bag. 14,602 total for all units 2013-06-30
2004-03-26 Not classified Other In some cases the device may develop small cracks that could expand over time and result in reduced shelf life or other failure. 60 units 2004-07-07
2004-03-26 Not classified Other In some cases the device may develop small cracks that could expand over time and result in reduced shelf life or other failure. 10 units 2004-07-07
2004-03-26 Not classified Other In some cases the device may develop small cracks that could expand over time and result in reduced shelf life or other failure. 3070 units 2004-07-07

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer