Avea 2 Ventilator
Avea Ventilator
Who Else Makes It
Concentrated in Guangdong (4), Jiangsu (2), Fujian (1). 6 of them are registered as contract manufacturers, able to build to a buyer’s own specification.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| GaleMed (Xiamen) Co., Ltd. | Xiamen | Fujian | Contract manufacturer | 33 |
| Vincent Medical (Dong Guan) Manufacturing Co., Ltd. | Dong Guan | Guangdong | Contract manufacturer | 29 |
| Sanmina | Kunshan | Jiangsu | Contract manufacturer | 15 |
| Mehow Innovative (Huizhou) Ltd. | Huizhou | Guangdong | Contract manufacturer | 11 |
| Respironics Medical Products (Shenzhen) Co., Ltd. | Shenzhen | Guangdong | Contract manufacturer | 8 |
| Beijing Aeonmed Co., Ltd. | Sanhe | Hebei | Contract manufacturer | 7 |
| Shenzhen Mindray BIO-Medical Electronics Co., Ltd. | Shenzhen | Guangdong | Specification developer | 61 |
| GE Medical Systems (China) Co., Ltd. | Wuxi | Jiangsu | Contract manufacturer | 57 |
Record
| Listed by | Vyaire Medical, Inc., United States |
|---|---|
| FDA product code | CBK |
| Device type | Ventilator, Continuous, Facility Use |
| Device class | Class 2 |
| Regulation | 21 CFR 868.5895 |
| 510(k) | K103211 |
| Record published | 2018-04-05 |
| Record revised | 2024-07-30 (version 5) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | serial number |
| Storage | Storage environment temperature -20 to 60 °C |
| Distribution end date | 2020-12-31 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 32029-001 | Withdrawn |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Puritan Bennett | Covidien LP | 285 |
| Newport | Newport Medical Instruments, Inc. | 103 |
| LTV | Vyaire Medical, Inc. | 92 |
| Ventilator, Continuous, Facility Use | Hans Rudolph, Inc. | 66 |
| Infant Flow SiPAP | Carefusion | 62 |
| Avea | Carefusion | 57 |
| Ventoux | Flight Medical Innovations Ltd. | 48 |
| Trilogy | Respironics Inc. | 42 |
| Avea™ | Vyaire Medical, Inc. | 39 |
| Patient Circuits | Zoll Medical Corporation | 38 |
| Performatrak | Respironics Inc. | 35 |
| Bio-Med Devices, Inc. | BIO-MED Devices, Inc. | 34 |
Showing 12 of 419. All products under Ventilator, Continuous, Facility Use
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K201082 | 2021-11-10 |
Recalls Naming This Product
15 recalls name this product by its own identifiers, the 510(k) it is cleared under.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|---|---|
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2016-05-17 | Class I | Under Investigation by firm | CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Mod | 922 units total (501 units in US) | 2017-09-28 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2015-04-03 | Class I | Process control | A pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing spontaneous breathing patients to breathe | 15,905 units total (5,670 units in US) | 2018-07-23 |
| Carefusion 211 Inc dba Carefusion | 510(k) number | 2013-09-05 | Class I | Component design/selection | CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a fiel | 19,546 units | 2015-06-04 |
| Carefusion 211 Inc | 510(k) number | 2011-09-08 | Not classified | Nonconforming Material/Component | The recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is voluntarily initiating | 6797 total, 1891 in US, 4906 worldwide | 2013-10-29 |
Recalls
1 of these is Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2023-05-05 | Class II | Nonconforming Material/Component | Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang | 910 units | |
| 2023-05-05 | Class II | Nonconforming Material/Component | Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang | 392 units | |
| 2021-12-20 | Class I | Software change control | Potential cessation in ventilation can occur under specific conditions. | 4189 units | 2024-08-28 |
| 2019-11-25 | Class II | Software design | The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a ‘no alarm’ condition, or presence of non-responsive touch screen. | 116 | 2023-08-08 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.