Trilogy 100

Trilogy 100 Ventilator, Bluetooth, Australia

Registry build 21.09.26
1manufacturer
3catalogue numbers in this listing
5recalls on this code

Who Else Makes It

Registered in Guangdong. One is registered as a contract manufacturer, able to build to a specification.

Chinese manufacturers registered for FDA product code NOU
Manufacturer City Province Registered as Device types
Mehow Innovative (Huizhou) Ltd. Huizhou Guangdong Contract manufacturer 11

Record

Regulatory record for Trilogy 100
Listed by Respironics Inc., United States
FDA product code NOU
Device type Continuous, Ventilator, Home Use
Device class Class 2
Regulation 21 CFR 868.5895
Establishment Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter
510(k) K083526
Record published 2017-09-22
Record revised 2021-03-05 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number
Storage Storage environment humidity 15 to 95 % RH; Storage environment temperature -4 to 140 °F
Distribution end date 2021-03-05

Catalogue Numbers

Catalogue numbers for Trilogy 100
Catalogue number Size Status
Trilogy 100 Ventilator, Bluetooth, Australia In distribution
Trilogy 100 Ventilator, Korea (Philips) Withdrawn
Trilogy 100 Ventilator, Korea (Vitalaire) Withdrawn

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Vivo 50 Breas Medical, Inc. 17
Patient Circuit Flight Medical Innovations Ltd. 13
Astral Resmed PTY Ltd. 12
Trilogy Evo Respironics Inc. 10
Trilogy Respironics Inc. 10
Care Orchestrator Respironics Inc. 10
Trilogy 200 Respironics Inc. 5
Continuous Ventilator Flight Medical Innovations Ltd. 3
Vocsn Roll Stand Ventec Life Systems, Inc. 2
Sensor Respironics Inc. 2
Astral Double Limb Kit Resmed PTY Ltd. 1
Trilogy EV300 Respironics Inc. 1

Showing 12 of 28. All products under Continuous, Ventilator, Home Use

Recalls Naming This Product

8 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of Trilogy 100
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Respironics Inc. 510(k) number 2022-12-07 Class I Process design The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which c 20,807
Respironics Inc. 510(k) number 2022-12-07 Class I Process design The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which c 9
Respironics Inc. 510(k) number 2021-06-14 Class I Device Design The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals. 255810
Philips Respironics, Inc. 510(k) number 2016-01-18 Class II Software design Software Issue 104,508 units 2017-05-26
Respironics Inc. 510(k) number 2014-02-11 Class I Component design/selection Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients. 572 2015-04-17
Respironics Inc. 510(k) number 2012-04-27 Class I Component design/selection The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply. 173 2012-11-01
Respironics Inc. 510(k) number 2011-10-11 Not classified Process design Trilogy 100, 200 and 202 units are being recalled due to low impeller press-force used during the manufacturing process. If the press force were low enough, theoretically the impeller could move up the shaft and come into contact with the 127 2012-11-06
Respironics Inc. 510(k) number 2010-06-10 Not classified Nonconforming Material/Component It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event. 23 units 2010-11-18

Recalls

2 of these are Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity
2024-07-16 Class I Device Design Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. 90,910 US; 56,273 ROW
2024-07-16 Class I Device Design Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls. 90,910 US; 56,273 ROW
2021-07-26 Class II Labeling design Two software issues have been identified related to pressure increase in the device: The first issue is described as “Infant/Pediatric EFS Calibration Pressure Increase” and the second issue is described as “Pressure Drift (Continuous Usage 18936 (US); 4659 (OUS)
2021-07-26 Class II Labeling design Two software issues have been identified related to pressure increase in the device: The first issue is described as “Infant/Pediatric EFS Calibration Pressure Increase” and the second issue is described as “Pressure Drift (Continuous Usage 18588 (US); 5662 (OUS)
2021-07-26 Class II Labeling design Two software issues have been identified related to pressure increase in the device: The first issue is described as “Infant/Pediatric EFS Calibration Pressure Increase” and the second issue is described as “Pressure Drift (Continuous Usage 101 (FEMA Stockpile Only); 21077 (OUS)

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer