| 2020-10-01 | Class II | Labeling Change Control | Dilation catheter packaged with the wrong size balloon. The labeling states that the balloon size is 6mm x 40mm, however included is a 5mm x 40mm balloon. Use of a balloon that has a smaller diameter than expected could result in inconseq | 281 units | 2022-05-25 |
| 2018-04-16 | Class II | Process control | Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endoto | 130 units | 2021-03-02 |
| 2018-04-16 | Class II | Process control | Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endoto | 140 units | 2021-03-02 |
| 2015-03-23 | Class II | Process control | Bard Peripheral Vascular is recalling the Bard ConQuest PTA Balloon Dilatation Catheter because it may be at risk of having deflation related issues. | 2620 units | 2015-08-28 |
| 2013-05-28 | Class II | Mixed-up of materials/components | Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm | 415 | 2013-10-23 |
| 2013-01-23 | Class II | Process control | Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker’s and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the s | 17 | 2013-06-10 |
| 2013-01-23 | Class II | Process control | Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker’s and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the s | 223 | 2013-06-10 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 213 | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 194 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 212 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 135 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 46 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 207 | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 522 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 449 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 72 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 250 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 56 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 227 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 96 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 406 units | 2012-02-01 |
| 2010-05-04 | Not classified | Nonconforming Material/Component | The affected DORADO PTA Balloon Dilatation catheters contain catheter materials that, under certain conditions, are subject to premature breakage allowing the balloon to partially or completely separate from the catheter shaft. | 194 units | 2012-02-01 |
| 2003-09-18 | Not classified | Other | Mislabeled. 7mm balloon labeled 12mm | 124 | 2004-12-10 |