CS100 Intra-Aortic Balloon Pump, Portuguese, 110v

Catalogue number 0998-00-3013-69, listed by Datascope Corp, United States, under System, Balloon, Intra-Aortic And Control. 1 manufacturer in China is registered for this product code.

Registry build 21.09.26
1manufacturer
1catalogue number
64510(k) clearances naming this product code
40recalls on this code

Who Else Makes It

Registered in Guangdong. One is registered as a contract manufacturer, able to build to a specification.

Chinese manufacturers registered for FDA product code DSP
Manufacturer City Province Registered as Device types
Amphenol Medical Technologies (Dongguan) Co., Ltd. Dongguan Guangdong Contract manufacturer 77

Record

Regulatory record for CS100 Intra-Aortic Balloon Pump, Portuguese, 110V
Listed by Datascope Corp, United States
FDA product code DSP
Device type System, Balloon, Intra-Aortic And Control
Device class Class 2
Regulation 21 CFR 870.3535
Establishment Contract manufacturer, Specification developer, Complaint-file establishment, Manufacturer
510(k) K031636
Record published 2016-09-21
Record revised 2023-03-02 (version 5)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Prescription Prescription only
Label carries serial number
Storage Storage environment atmospheric pressure 483 to 760 Millimeter of Mercury; Handling environment humidity 5 to 95 % RH; Storage environment temperature -20 to 70 °C
Distribution end date 2016-12-31

Catalogue Numbers

Catalogue numbers for CS100 Intra-Aortic Balloon Pump, Portuguese, 110V
Catalogue number Size Status
0998-00-3013-69 Withdrawn

Same Product Code

Showing 12 of 268. All products under System, Balloon, Intra-Aortic And Control

510(K) Clearances

FDA 510(k) clearances naming this product code
Clearance Decision date
K181122 2018-05-31
K172305 2017-10-12
K163542 2017-01-31
K151254 2015-07-02
K133074 2013-12-12
K063525 2007-01-05
K041281 2004-06-07
K031636 2003-08-11
K031569 2003-06-10
K020257 2002-04-24
K013326 2001-11-02
K003598 2000-12-21
K002949 2000-10-20
K994157 2000-03-13
K980780 1998-05-29
K980385 1998-05-01
K965236 1997-07-01
K964987 1997-03-17
K961509 1996-11-15
K960166 1996-04-24
K942454 1995-02-24
K944460 1994-12-09
K940231 1994-09-13
K940178 1994-08-09
K911000 1993-08-25
K914802 1993-08-25
K920427 1993-08-03
K905663 1993-08-02
K922235 1993-04-27
K921638 1993-02-10
K913847 1992-03-11
K912629 1992-03-11
K910997 1991-11-05
K905056 1991-06-11
K903930 1990-11-16
K904414 1990-10-22
K904018 1990-10-19
K903769 1990-10-19
K903867 1990-10-19
K903893 1990-10-19
K903892 1990-10-19
K903736 1990-10-09
K902057 1990-09-25
K896497 1990-08-17
K896430 1990-02-05
K893243 1989-06-20
K880289 1988-06-10
K863219 1987-01-14
K850482 1985-08-22
K833828 1985-02-08
K831235 1985-02-08
K831397 1983-11-02
K831307 1983-11-02
K823342 1983-01-24
K821951 1982-12-21
K820882 1982-10-18
K821065 1982-08-19
K813182 1982-03-31
K811372 1981-08-03
K810297 1981-05-06
K802167 1981-01-13
K801601 1981-01-13
K790775 1979-10-01
K780280 1978-03-15

Recalls Naming This Product

7 recalls name this product by its own identifiers, the 510(k) it is cleared under.

Recalls naming the identifiers of CS100 Intra-Aortic Balloon Pump, Portuguese, 110V
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Datascope Corp 510(k) number 2026-01-23 Class II Under Investigation by firm The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated. 10,897 units
Datascope Corp 510(k) number 2017-07-17 Class I Under Investigation by firm False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy. 14,547 units (5,049 domestically & 9,498 units international
Datascope Corp 510(k) number 2017-07-17 Class I Under Investigation by firm False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy. 14,547 units (5,049 domestically & 9,498 units international
Datascope Corp 510(k) number 2017-06-16 Class I Component design/selection The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy. 12,319 units total 2019-11-29
Datascope Corp 510(k) number 2017-06-16 Class I Component design/selection The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy. 12,319 units total 2019-11-29
Datascope Corp 510(k) number 2014-03-21 Class I Nonconforming Material/Component Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning. 11,881 units (US 5183 and OUS 6698) 2018-03-20
Datascope Corp 510(k) number 2010-05-10 Not classified Other Certain MAQUET/Datascope Intra-Aortic Balloon Pumps (IABP) have the potential to experience display related issues. Although the IABP continues to deliver therapy to the patient, display related issues may cause the user to be unable to vie 161 total devices 2010-10-27

Recalls

16 of these are Class I. Class I is the FDA’s most serious grade: a reasonable probability of serious injury or death. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity
2026-02-06 Class III Under Investigation by firm The Battery Charging Station is an optional dual-bay
charger used to charge the Cardiosave IABP Lithium-
Ion batteries when they are not being used to
operate the IABP. Getinge has identified that a
protruding screw in the left battery bay
28 units (20 US, 8 OUS)
2026-01-23 Class II Under Investigation by firm The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated. 10,897 units
2025-11-04 Class II Component design/selection The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023. 11,470 units
2025-11-04 Class II Component design/selection The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023. 47 units
2025-10-23 Class III Process change control The IFU addendum updates the Vibration and Shock Table to reference the correct standards. 47 units (44 US, 3 OUS)
2025-10-23 Class III Process change control The IFU addendum updates the Vibration and Shock Table to reference the correct standards. 11470 units (5095 US, 6375 OUS)
2024-09-17 Class II Error in labeling There is discrepant labeling on the inner and outer packaging of the device. 160 units
2024-08-08 Class I Software design Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-A 5,475 units
2024-08-08 Class I Software design Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Int 5,475 units
2024-08-08 Class I Software design Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-A 5,475 units
2024-08-08 Class I Software design Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Int 5,475 units
2023-07-31 Class II Under Investigation by firm Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector. 9379 units
2023-07-31 Class II Under Investigation by firm Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector. 9379 units
2023-07-31 Class I Device Design Users reported “System Over Temperature” alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode. 11574 units (5413 US, 6161 OUS)
2023-07-31 Class I Under Investigation by firm Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the 9175 units
2023-07-31 Class I Under Investigation by firm Users were identifying autofill failure conditions on the devices causing pump stops. 9175 units
2023-07-31 Class I Under Investigation by firm Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the 9175 units
2023-07-31 Class I Under Investigation by firm Users were reporting instances of “Gas Loss in IAB Circuit” and “Gas Gain in IAB Circuit” alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not 9175 units
2023-07-31 Class I Under Investigation by firm Users were reporting instances of “Gas Loss in IAB Circuit” and “Gas Gain in IAB Circuit” alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not 9175 units
2023-07-31 Class I Under Investigation by firm Users were identifying autofill failure conditions on the devices causing pump stops. 9175 units
2023-07-31 Class I Device Design Users reported “System Over Temperature” alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode. 47 units (42 US, 5 OUS)
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-23 Class II Under Investigation by firm During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; su 321,609 total kits
2023-06-05 Class I Device Design IABP may lose the ability to charge batteries in one or both bay slots.

Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC
power. Low battery alarms may alert the User to the issue prior to interr

9174 (4586 US; 4588 OUS)
2023-06-05 Class I Device Design IABP may lose the ability to charge batteries in one or both bay slots.

Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC
power. Low battery alarms may alert the User to the issue prior to interr

9174 (4586 US; 4588 OUS)
2023-06-05 Class I Component design/selection An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the
Power Management Board and/or Solenoid Board.
This issue may lead to an unexpected interruption of therapy.
9174 (4586 US; 4588 OUS) in total
2023-06-05 Class I Component design/selection An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the
Power Management Board and/or Solenoid Board.
This issue may lead to an unexpected interruption of therapy.
9174 (4586 US; 4588 OUS) in total
2023-05-04 Class III Under Investigation by firm The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU. 4 units
2023-04-04 Class II Component design/selection The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle. 54 units
2023-02-28 Class II Under Investigation by firm There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB i 391 total IABPs
2023-02-28 Class II Under Investigation by firm There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume.

This potential failure of an impacted Safety Disk may lead to slightly reduced IAB i

391 total IABPs
2023-02-28 Class II Under Investigation by firm There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB i 669 safety disks
2023-02-28 Class II Under Investigation by firm There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB i 10

Source: U.S. FDA establishment registration and device listing records, public domain.

Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker
31provinces with a registered manufacturer