| Maquet Medical Systems, Usa | 510(k) number | 2024-01-16 | Class II | Nonconforming Material/Component | The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection. | 16520 worldwide; 6657 US | |
| Maquet Medical Systems, Usa | 510(k) number | 2024-01-16 | Class II | Nonconforming Material/Component | The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection. | 16520 worldwide; 6657 US | |
| Maquet Medical Systems, Usa | 510(k) number | 2023-02-13 | Class II | Process control | Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia. | 10,587 units (US); 94,987 (worldwide) | |
| Maquet Medical Systems, Usa | 510(k) number | 2022-07-22 | Class II | Package design/selection | The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020). | 206 US; 48,960 Worldwide | |
| Maquet Medical Systems, Usa | 510(k) number | 2022-07-22 | Class II | Package design/selection | The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020). | 206 US; 48,960 Worldwide | |
| Maquet Medical Systems, Usa | 510(k) number | 2022-07-22 | Class II | Package design/selection | The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020). | 6,871 US; 18,670 worldwide | |
| Maquet Medical Systems, Usa | 510(k) number | 2022-07-22 | Class II | Package design/selection | The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020). | 6,871 US; 18,670 worldwide | |
| Maquet Cardiovascular Us Sales, Llc | 510(k) number | 2020-11-20 | Class II | Package design/selection | The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product’s sterility barrier. | 58,438 devices worldwide | 2024-11-07 |
| Maquet Cardiovascular Us Sales, Llc | 510(k) number | 2020-11-20 | Class II | Package design/selection | The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product’s sterility barrier. | | 2024-11-07 |
| Maquet Cardiovascular Us Sales, Llc | 510(k) number | 2019-07-11 | Class II | Under Investigation by firm | The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly | 54 | 2020-12-11 |
| Maquet Cardiovascular Us Sales, Llc | 510(k) number | 2013-08-15 | Class II | Nonconforming Material/Component | The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet con | 1,182 units | 2017-07-11 |
| Maquet Cardiovascular Us Sales, Llc | 510(k) number | 2012-06-12 | Class II | Nonconforming Material/Component | It has come to firm’s attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped. | 23,203 units | 2017-09-29 |