Capillary Adapter Cleaning Kit
EasyLyte Capillary Adapter Cleaning Kit
2manufacturers in China registered for this device type
1catalogue number
4510(k) clearances naming this product code
1recall
Who Else Makes It
All of them in Guangdong.
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Shenzhen Mindray BIO-Medical Electronics Co., Ltd. | Shenzhen | Guangdong | Specification developer | 61 |
| Edan Instruments, Inc. | Shenzhen | Guangdong | Contract manufacturer | 56 |
Record
| Listed by | Medica Corp., United States |
|---|---|
| FDA product code | CEM |
| Device type | Electrode, Ion Specific, Potassium |
| Device class | Class 2 |
| Regulation | 21 CFR 862.1600 |
| Establishment | Repacker / relabeller, Manufacturer |
| 510(k) | K864872 |
| Record published | 2018-12-04 |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Label carries | lot or batch number, expiry date, manufacturing date |
| Storage | Storage environment temperature 4 to 25 °C |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 2292 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Nova | Nova Biomedical Corporation | 8 |
| Stat Profile | Nova Biomedical Corporation | 7 |
| Potassium electrode | Roche Diagnostics Gmbh | 4 |
| Potassium Electrode | Diamond Diagnostics Inc. | 3 |
| Synermed® | Infrared Laboratory Systems LLC | 3 |
| Architect | Abbott Laboratories | 3 |
| Phoenix Diagnostics, Inc. | Phoenix Diagnostics, Inc. | 2 |
| I.S.E. K | Randox Laboratories Limited | 2 |
| Pointe Scientific, Inc. | Horiba Instruments Incorporated | 2 |
| Potassium Electrode | Medica Corp. | 2 |
| Na/K Solution Pack | Medica Corp. | 2 |
| Selectra | VitalScientific B.V. | 2 |
Showing 12 of 93. All products under Electrode, Ion Specific, Potassium
510(K) Clearances
| Clearance | Decision date |
|---|---|
| K220396 | 2023-09-26 |
| K112788 | 2011-10-20 |
| K961326 | 1996-06-21 |
| K844044 | 1984-11-27 |
Recalls
1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|
| 2025-12-31 | Class II | Nonconforming Material/Component | A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes. | 1,578 units |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.