Vitros
VITROS Chemistry Products LIPA Slides
Who Else Makes It
Concentrated in Jilin (1), Shandong (1).
| Manufacturer | City | Province | Registered as | Device types |
|---|---|---|---|---|
| Dirui Industrial Co., Ltd. | Changchun | Jilin | Contract manufacturer | 84 |
| Biobase Biodustry(Shandong)Co., Ltd. | Jinan | Shandong | Manufacturer | 70 |
Record
| Listed by | Ortho-Clinical Diagnostics, Inc., United States |
|---|---|
| FDA product code | CHI |
| Device type | Lipase-Esterase, Enzymatic, Photometric, Lipase |
| Device class | Class 1 |
| Regulation | 21 CFR 862.1465 |
| Establishment | Contract manufacturer, Specification developer, Repacker / relabeller, Complaint-file establishment, Manufacturer, Exporter |
| Premarket notification | Exempt |
| Record published | 2020-04-18 |
| Record revised | 2023-04-26 (version 2) |
Specifications
| MRI safety | Labeling does not contain MRI Safety Information |
|---|---|
| Prescription | Prescription only |
| Supplied as a kit | Yes |
| Label carries | lot or batch number, expiry date |
| Storage | Storage environment temperature -18 to -18 °C |
| Packaging | Case of 30 |
Catalogue Numbers
| Catalogue number | Size | Status |
|---|---|---|
| 1668409 | In distribution | |
| 8297749 | In distribution |
Same Product Code
| Product | Sold by | Catalogue numbers |
|---|---|---|
| Biobase | Biobase Biodustry(Shandong)Co., Ltd. | 7 |
| Lipase (LPS) | Randox Laboratories Limited | 4 |
| Carolina Liquid Chemistries Corp.® Lipase Reagent KIT ST621 | Carolina Liquid Chemistries | 4 |
| Lipase colorimetric assay | Roche Diagnostics Gmbh | 3 |
| Architect | Abbott Ireland | 2 |
| Pointe Scientific, Inc. Lipase-Color | Horiba Instruments Incorporated | 2 |
| Alinity | Abbott Ireland | 2 |
| Architect | Abbott Laboratories | 1 |
| Dimension Vista® Flex® reagent cartridge Lipl | Siemens Healthcare Diagnostics Inc | 1 |
| MedTest Dx, Inc. Lipase (Colorimetric) | Medtest DX | 1 |
| Lipase NG OC Reagent Kit | Sentinel CH. SPA | 1 |
| Lipase colorimetric | Roche Diagnostics Gmbh | 1 |
Showing 12 of 28. All products under Lipase-Esterase, Enzymatic, Photometric, Lipase
Recalls Naming This Product
1 recall names this product by its own identifier, the UDI on its label.
| Recalled by | Identified by | Initiated | FDA class | Root cause | Reason | Quantity |
|---|---|---|---|---|---|---|
| Ortho-Clinical Diagnostics, Inc. | UDI-DI | 2023-09-12 | Class II | Under Investigation by firm | Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis o | 342 units |
Recalls
1 of these is Class II. Class II means a temporary or reversible health consequence. Volume alone is not a quality signal: a manufacturer that ships more recalls more.
| Initiated | FDA class | Root cause | Reason | Quantity | Closed |
|---|---|---|---|---|---|
| 2013-06-12 | Class II | Equipment maintenance | Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues. | 62,039 units (26,765 units domestically & 35,274 units inter | 2017-04-11 |
| 2006-10-09 | Not classified | Other | The firm identified circumstances in which biased results or calibration failures can be observed when using any lot of VITROS LIPA slides if they have been stored on the VITROS Chemistry Systems for more than 2 days. | Cat./REF # 166 8409: Domestic – 3031, Foreign – 2511; Cat./ | 2008-05-22 |
Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.