QC Sets and Panels: Kwik-Stik™

QC Sets and Panels: KWIK-STIK™ Enteric Bacterial Verification Panel (Live Culture)

Registry build 21.09.26
2manufacturers in China registered for this device type
7catalogue numbers in this listing
19recalls on this code

Who Else Makes It

Concentrated in Jilin (1), Guangdong (1).

Chinese manufacturers registered for FDA product code OHQ
Manufacturer City Province Registered as Device types
Dirui Industrial Co., Ltd. Changchun Jilin Contract manufacturer 84
Phase Scientific (Shenzhen) Company Limited Shenzhen Guangdong Repacker / relabeller 8

Record

Regulatory record for QC Sets and Panels: Kwik-Stik™
Listed by Microbiologics, Inc., United States
FDA product code OHQ
Device type Multi-Analyte Controls Unassayed
Device class Class 1
Regulation 21 CFR 862.1660
Establishment Contract manufacturer, Repacker / relabeller, Manufacturer
Premarket notification Exempt
Record published 2020-04-30
Record revised 2022-09-02 (version 2)

Specifications

Specifications
MRI safety Labeling does not contain MRI Safety Information
Single use Yes
Supplied as a kit Yes
Label carries lot or batch number, expiry date
Storage Storage environment temperature 2 to 8 °C

Catalogue Numbers

Catalogue numbers for QC Sets and Panels: Kwik-Stik™
Catalogue number Size Status
8179 In distribution
8178 In distribution
8174 In distribution
8173 In distribution
8171 In distribution
8243 In distribution
8164 In distribution

Same Product Code

Other products under the same FDA product code
Product Sold by Catalogue numbers
Exact Diagnostics Exact Diagnostics 64
NATtrol™ Zeptometrix LLC. 54
QC Sets and Panels: Helix Elite™ Microbiologics, Inc. 51
Helix Elite™ Molecular Standards: Inactivated Standard Microbiologics, Inc. 41
Multichem S Plus Techno-Path Manufacturing Ltd. 35
Virotrol Bio-Rad Laboratories Inc. 27
Helix Elite™ Molecular Standards: Synthetic Standard Microbiologics, Inc. 25
Multichem IA Plus Techno-Path Manufacturing Ltd. 17
D3 Diagnostic Hybrids, Inc. 14
Liquichek Bio-Rad Laboratories Inc. 10
Multichem IA Techno-Path Manufacturing Ltd. 8
QC Sets and Panels: Kwik-Stik™ 2 Pack Microbiologics, Inc. 8

Showing 12 of 107. All products under Multi-Analyte Controls Unassayed

Recalls Naming This Product

2 recalls name this product by its own identifiers, the UDI on its label.

Recalls naming the identifiers of QC Sets and Panels: Kwik-Stik™
Recalled by Identified by Initiated FDA class Root cause Reason Quantity Closed
Microbiologics, Inc. UDI-DI 2020-12-16 Class III Employee error The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have been 7/31/2022. 2023-01-12
Microbiologics, Inc. UDI-DI 2015-12-01 Class II Material/Component Contamination The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This caus 541 units 2022-07-07

Recalls

18 of these are Class II. Class II means a temporary or reversible health consequence. Class III is unlikely to cause harm at all → typically a labelling or documentation fault. Volume alone is not a quality signal: a manufacturer that ships more recalls more.

Recalls referencing this product code
Initiated FDA class Root cause Reason Quantity Closed
2025-02-20 Class II Nonconforming Material/Component The RSV target may give a late Ct value and could potentially not pass QC. 28 units
2025-02-20 Class II Process control The A549 human cell target (human cells for sample adequacy control) may give a late Ct value. 6 units
2025-01-20 Class II Under Investigation by firm Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 27 units
2025-01-20 Class II Under Investigation by firm Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 16 units
2024-01-08 Class II Process control The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control. 580 packages
2022-07-27 Class III Process control Product not registered for use in the UK 2 2024-08-02
2022-03-24 Class II Process control QC process was not adequate for the specification range. 1 ea 2023-01-09
2022-03-22 Class II Process control The QC process was not adequate for the specification range. 2 ea 2023-01-09
2022-03-03 Class II Incorrect or no expiration date Shelf-life for the 8244 Pneumonia (33 Targets) Control Panel determined to be 6 months instead of 18 months. 13 units 2024-01-16
2022-02-25 Class II Nonconforming Material/Component Distributed product did not undergo proper release testing. 80 units 2024-01-16
2021-08-20 Class II Material/Component Contamination The product indicated is contaminated with a plasmid DNA that contains the target SARS-CoV-2 sequences. 127 2023-01-12
2021-08-20 Class II Material/Component Contamination The SARS-CoV-2 Process Control (Pellet) contains RNA transcripts that include diagnostically relevant SARS-CoV-2 targets (CDC and WHO consensus sequences). 218 2023-01-12
2020-01-03 Class II Process control Product 8208 Vaginal Verification Panel is a 6 pool kit and has contaminant Candida krusei in Pool 2 (subcomponent 6063). Candida krusei is an organism intended to be contained in Pool 6 of the kit. This should have no impact on patient te 9 2020-09-16
2019-12-16 Class II Device Design The recommended hydration fluid for the 8180 Gram-Positive Blood Culture Control Panel (Inactivated Pellet), has changed from sterile 0.85% saline or sterile blood culture media to sterile DI water or pH 7.2 phosphate buffer. 810 2020-12-11
2017-06-06 Class II Process design Real-time shelf life testing failed at 24 months 1 2023-01-17
2017-06-06 Class II Process design Real-time shelf life testing failed at 24 months 29 units 2023-01-17
2017-06-06 Class II Process design Real-time shelf life testing failed at 24 months 1 unit 2023-01-17
2017-06-06 Class II Process design Real-time shelf life testing failed at 24 months 1 unit 2023-01-17
2017-06-06 Class II Process design Real-time shelf life testing failed at 24 months 8 units 2023-01-17

Source: U.S. FDA establishment registration and device listing records, public domain. Restructured for clarity and searchability. Not affiliated with or endorsed by the FDA. Records are records: nothing here is an offer of sale.

Where China Actually Supplies

Every device page names the company that lists it and the factories registered to make the same thing. These are the sections with the most of them.

5,100manufacturers registered in China
2,581device types they are registered for
340,248device records, each naming its maker